Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1341-2024

MINI STICK MAX 5F X 10 CM STIFF .018 NI/TU ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-759

Publicado el 27 de marzo de 2024 · Inicio del retiro: 30 de enero de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Angiodynamics, Inc.
Distribuidor
Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.
Lote
UDI: 15051684023023 UPN: H965457591 Lot Number: 5789493 5789494 5789495 5789496 5789497 5791596 5791597 5791598 5793392 5793393 5794602 5794603 5794604 5794605 5794606 5795874 5795875 5797457 5797458 5797466 5797467 5798837 5798843 5798844
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States