Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1396-2013

Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Reamer, 6", Part Number 2110-0504 Product Usage: Facilitate placement of bone cement in weakened or diminished bone

Publicado el 5 de junio de 2013 · Inicio del retiro: 10 de noviembre de 2012

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Orthovita, Inc., dBA Stryker Orthobiologics.
Distribuidor
USA Nationwide Distribution in the states of AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, ND, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, and WI.
Lote
Manufacturer Part Number 2110-0504, Lot Number E909007R
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta