Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1776-2025

Intera 1.5T Master/Nova, Model Number: 781106;

Publicado el 21 de mayo de 2025 · Inicio del retiro: 24 de marzo de 2025

Resumen

FDA registra el retiro Z-1776-2025 de dispositivos médicos de Clase II, publicado el 21 de mayo de 2025 e iniciado el 24 de marzo de 2025. Empresa responsable: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Estado: En curso; Cantidad reportada: 103 units (37 US, 66 OUS); Lote: Model Number (REF): 781106; UDI-DI: N/A; All serial numbers. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Distribuidor
US: Nationwide Distribution International: Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Botswana, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam, Yemen, Zambia.
Lote
Model Number (REF): 781106; UDI-DI: N/A; All serial numbers.
Clasificación de riesgo
Class II
Área (FDA)
device
País
Netherlands

Retiros relacionados

Preguntas sobre este retiro

¿Intera 1.5T Master/Nova, Model Number: 781106; fue retirado del mercado?

Sí. FDA publicó el retiro Z-1776-2025 el 21 de mayo de 2025. Peligro reportado: The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter