Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1778-2017

Merge Hemo software. The firm name on the labeling is Merge Healthcare, Hartland, WI. Merge Hemo monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure

Publicado el 19 de abril de 2017 · Inicio del retiro: 4 de abril de 2016

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Use of the software may show an incorrect value to the user when viewing the Fractional Flow Reserve (FFR) results during recording.

Qué hacer

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Detalles del retiro

Fabricante
Merge Healthcare, Inc.
Distribuidor
Nationwide Distribution including AZ, CA, CO, FL, GA, KS, LA, MI, MO, OH, OK, TN, TX, VA, WA, WI, and WV.
Lote
Versions 9.10.0, 9.20.0, 9.20.1, 9.20.2, 9.30.3, and 9.40.0.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta