PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT

Publicado el 6 de mayo de 2020 · Inicio del retiro: 11 de marzo de 2020

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Merit Medical Systems, Inc.
Distribuidor
U.S.: NV, IL, VA, PA, CA, AR, MA, GA, OH, NJ, TX, FL, NC, SC, MI, NH, CO, AL, MN, KS, NM, AZ, ID, MD, NY, OR, DC, DE, MO, IN, WI, IA, TN. O.U.S.: Switzerland, Hong Kong, Spain, Portugal, Canada, Singapore, United Arab Emirates, Mexico, United Kingdom, France, Japan, Germany, Italy
Lote
SDRB-REG-LT: Lot(UDI): H1559517(10884450363673) and H1727668(00884450363676). SDRB-REG-RT: Lot(UDI): H1666548(10884450363659) and H1723973(10884450363659)
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States