Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system

Publicado el 27 de junio de 2012 · Inicio del retiro: 11 de mayo de 2012

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Philips Healthcare Inc.
Distribuidor
6 US customers
Lote
All systems with Eleva software version 4.x and stitching option.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States