FDAFDA · Alimentos y medicamentos·N.º oficial Z-1823-2016

Various trauma and sports medicine instruments and implants. Screw, fixation, bone.

Publicado el 1 de junio de 2016 · Inicio del retiro: 1 de abril de 2016

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an infection.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Biomet, Inc.
Distribuidor
Worldwide Distribution - US (Nationwide) and globally to Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecaudor, The Netherlands, Japan, Korea, Malaysia, Mexico, Peru.
Lote
Part #'s HAA60-11030 HAA60-13030 HAA60-15040 HAA60-16040 HAA60-18040 Lot #'s 016150 016210 165130 016230 133890 181600 133970 134000 181620
Clasificación de riesgo
Class II