Eon Mini - 3788 (IPG), Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back, surgery syndrome, intractable low back and leg pain.
Publicado el 7 de agosto de 2013 · Inicio del retiro: 26 de julio de 2012
Resumen
FDA registra el retiro Z-1836-2013 de dispositivos médicos de Clase II, publicado el 7 de agosto de 2013 e iniciado el 26 de julio de 2012. Empresa responsable: St. Jude Medical; Estado: Finalizado; Cantidad reportada: 34,617; Lote: Multiple lots recalled are the ones with supplier's inner battery lots of 326-508. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
As part of St. Jude Medical's routine tracking and product monitoring, they have received a total of 214 reports, out of 34,617 Eon Mini IPGs, that lost the ability to communicate or recharge due to an inner battery weld issue resulting in loss of pain relief and subsequent explant. Explant surgery, as with any surgery, presents a risk to patient health. Adverse events associated with an unplanne
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- St. Jude Medical
- Distribuidor
- Worldwide Distribution--USA (nationwide) including Puerto Rico and the countries of Argentina, Australia, Austria, Belgium, Brazil, canada, Colombia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Ireland, Israel, Italy, Japan, Mexico, Netherlands, Saudi, South Africa, Spain, Sweden, Switzerland, Turkey and United Kingdom.
- Lote
- Multiple lots recalled are the ones with supplier's inner battery lots of 326-508
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACardioMEMS Heart Failure (HF) System that consist of: -CardioMEMS Hospital System Model CM3100 -CardioMEMS2 de abril de 2025
- FDASt. Jude Medical Agilis NxT Steerable Introducer, REF 40830924 de julio de 2024
- FDACardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides15 de marzo de 2023
- FDACardioMEMS HF System PA Sensor and Delivery System, Model Number CM2000, The CardioMEMS HF System provides15 de marzo de 2023
- FDAAU5800 Clinical Chemistry Analyzer with Automation Ready Connection (All Software Versions), Model Numbers7 de agosto de 2013
- FDAOmnicycle Elite; US Importer and Distributor: Accelerated Care Plus, Reno NV. Manufactured by Medica7 de agosto de 2013
Preguntas sobre este retiro
¿Eon Mini - 3788 (IPG), Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back, surgery syndrome, intractable low back and leg pain. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1836-2013 el 7 de agosto de 2013. Peligro reportado: As part of St. Jude Medical's routine tracking and product monitoring, they have received a total of 214 reports, out of 34,617 Eon Mini IPGs, that lost the ability to communicate or recharge due to an inner battery weld issue resulting in loss of pain relief and subsequent explant. Explant surgery, as with any surgery, presents a risk to patient health. Adverse events associated with an unplanne
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter