Alere Triage BNP PN 98000XR Intended to be used as an aid in the diagnosis and assessment of severity of congestive heart failure (also referred to as heart failure).

Publicado el 18 de julio de 2012 · Inicio del retiro: 22 de mayo de 2012

Resumen

FDA registra el retiro Z-1931-2012 de dispositivos médicos de Clase I, publicado el 18 de julio de 2012 e iniciado el 22 de mayo de 2012. Empresa responsable: Alere San Diego, Inc.; Estado: Finalizado; Cantidad reportada: 30,083 kits; Lote: Phase 1: K50671B, W49486B, W49487B, W49489B, W49498B, W49500B, W49873B, W49890B, W49895B, W49896B, W49934B, W49947B, W49948B, W49950B, W49962B, W49968B, W50257B, W50278B, W50279B, W50282B, W50287B, W50294B, and W50313B. Phase 2: K50665B, W49855B, W49862B, W49865B, W49887B, W49892B, W49893B, W49894B, W49900B, W49933B, W49939B, W49953B, W49958B, W49960B, W49961B, W50252B, W50268B, W50273B, W50274B, W50290B, W50292B, W50295B, W50297B, W50298B, W50304B, W50355B, W50357B, and W50375B. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Alere San Diego, Inc.
Distribuidor
Nationwide Distribution
Lote
Phase 1: K50671B, W49486B, W49487B, W49489B, W49498B, W49500B, W49873B, W49890B, W49895B, W49896B, W49934B, W49947B, W49948B, W49950B, W49962B, W49968B, W50257B, W50278B, W50279B, W50282B, W50287B, W50294B, and W50313B. Phase 2: K50665B, W49855B, W49862B, W49865B, W49887B, W49892B, W49893B, W49894B, W49900B, W49933B, W49939B, W49953B, W49958B, W49960B, W49961B, W50252B, W50268B, W50273B, W50274B, W50290B, W50292B, W50295B, W50297B, W50298B, W50304B, W50355B, W50357B, and W50375B.
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Alere Triage BNP PN 98000XR Intended to be used as an aid in the diagnosis and assessment of severity of congestive heart failure (also referred to as heart failure). fue retirado del mercado?

Sí. FDA publicó el retiro Z-1931-2012 el 18 de julio de 2012. Peligro reportado: Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter