CELLTRACKS ANALYZER II (Software Version 2.5.1 & Below. Product Usage: The CELLTRACKS AUTOPREP System is a general purpose laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood. Cell analyzers such as the CellTracks Analyzer II, CellSpotter System, flow cytometers or microscopes may be used for cell identification and enumeration. The system is fit for in vitro diagnostic use. In the CellTracks Analyzer II software 2.5.1 release notes Pub. No. J40136EN issued 2012-03-26 page 1, the document states that the edit patient sample information screen now includes a patient information section, a new feature unique to software 2.5.1. This section provides a drop down for patient race and cancer type. There are also separate fields for first and last name for both the physician and patient. There is no mention of the drop down patient ID box on the right side of the user interface which auto-populates previous sample data.
Publicado el 21 de agosto de 2013 · Inicio del retiro: 28 de diciembre de 2012
Resumen
FDA registra el retiro Z-1950-2013 de dispositivos médicos de Clase II, publicado el 21 de agosto de 2013 e iniciado el 28 de diciembre de 2012. Empresa responsable: Veridex, LLC; Estado: Finalizado; Cantidad reportada: US: 70 Units; Ex-US 76 Units; Lote: CELLTRACKS ANALYZER II (Software Version 2.5.1 & Below). Product Code: CT1476B (9555). La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another patient's identification (PID) number. The purpose of this notification is to inform you of the issue and provide instructions to help prevent an occurrence on your CELLTRACKS AUTOPREP System and CELLTRACKS ANALYZER II.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Veridex, LLC
- Distribuidor
- Worldwide Distribution - USA Nationwide and the countries of Australia, Canada, China, India, Japan, Singapore, England, France, Germany, Italy and Spain.
- Lote
- CELLTRACKS ANALYZER II (Software Version 2.5.1 & Below). Product Code: CT1476B (9555).
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACELLTRACKS ANALYZER II® (Product Code 9555) Product Usage: The CELLTRACKS ANALYZER II® is a semi-automated19 de marzo de 2014
- FDACELLTRACKS AUTOPREP System (Software Version 2.4.1 & Below. Product Usage: The CELLTRACKS AUTOPREP System is a21 de agosto de 2013
- FDAISE Electrolyte Reference reagent (for Synchron LX Systems and UniCel DxC 600/800 Systems), Part No. A28937,21 de agosto de 2013
- FDAColorimetric (CM)/ Rate (RT) incubator wear pads. Product Codes: 1) 6801375, 2) 6801890, 3) 6802445, 4)21 de agosto de 2013
- FDASynthes Flexible Medullary Reamer. Intended to be used to facilitate the preparation of the intramedullary21 de agosto de 2013
- FDA1) Adult Manual Resuscitator, 2K8004, Oxygen Reservoir Bag, Adult Mask; 2) Adult Manual Resuscitator, 2K8005,21 de agosto de 2013
Preguntas sobre este retiro
¿CELLTRACKS ANALYZER II (Software Version 2.5.1 & Below. Product Usage: The CELLTRACKS AUTOPREP System is a general purpose laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood. Cell analyzers such as the CellTracks Analyzer II, CellSpotter System, flow cytometers or microscopes may be used for cell identification and enumeration. The system is fit for in vitro diagnostic use. In the CellTracks Analyzer II software 2.5.1 release notes Pub. No. J40136EN issued 2012-03-26 page 1, the document states that the edit patient sample information screen now includes a patient information section, a new feature unique to software 2.5.1. This section provides a drop down for patient race and cancer type. There are also separate fields for first and last name for both the physician and patient. There is no mention of the drop down patient ID box on the right side of the user interface which auto-populates previous sample data. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1950-2013 el 21 de agosto de 2013. Peligro reportado: Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another patient's identification (PID) number. The purpose of this notification is to inform you of the issue and provide instructions to help prevent an occurrence on your CELLTRACKS AUTOPREP System and CELLTRACKS ANALYZER II.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter