CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit

Publicado el 2 de julio de 2025 · Inicio del retiro: 28 de mayo de 2025

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Terumo Cardiovascular Systems Corporation
Distribuidor
US Nationwide distribution in the states of California, Delaware, Hawaii, Iowa, Indiana, Michigan, North Carolina, New York, Ohio, and Washington. The country of New Zealand.
Lote
UDI-DI 00886799001889 Lot Number/Serial Number Range: B0050231-B0050240 B0050221-B0050230 B0050211-B0050219 B0050201-B0050209 B0050191-B0050200
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States