CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit
Publicado el 2 de julio de 2025 · Inicio del retiro: 28 de mayo de 2025
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Terumo Cardiovascular Systems Corporation
- Distribuidor
- US Nationwide distribution in the states of California, Delaware, Hawaii, Iowa, Indiana, Michigan, North Carolina, New York, Ohio, and Washington. The country of New Zealand.
- Lote
- UDI-DI 00886799001889 Lot Number/Serial Number Range: B0050231-B0050240 B0050221-B0050230 B0050211-B0050219 B0050201-B0050209 B0050191-B0050200
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States