LMA Esophageal Intubation Detector (EID), Catalog Number EID100 The device is used to assist verification of placement of the endotracheal tube or esophageal/tracheal double lumen tube. This device is to be used as an adjunct to assess intubation. Its purpose is not to eliminate clinical judgment
Publicado el 25 de julio de 2012 · Inicio del retiro: 23 de mayo de 2012
Resumen
FDA registra el retiro Z-2022-2012 de dispositivos médicos de Clase II, publicado el 25 de julio de 2012 e iniciado el 23 de mayo de 2012. Empresa responsable: LMA North America Inc; Estado: Finalizado; Cantidad reportada: 18,799 units; Lote: 2009 Lot # 90923A, 91222A. 2010 Lot# 00327A, 00518A, 00623A, 00831B, 01013A, 01103A, 01214A. 2011 Lot# 10110A, 10301B, 10423A, 10530A, 10602A, 10703A, 10804A, 11004A. 2012 Lot# 120104, 120414. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The recall was initiated because LMANA has received a complaint that the Esophageal Intubation Detector (EID100) has a potential to create a false positive if used on a patient during confirmation of intubation.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- LMA North America Inc
- Distribuidor
- Worldwide Distribution--USA (nationwide) and the countries of Canada, United Kingdom, ZA, and Australia.
- Lote
- 2009 Lot # 90923A, 91222A. 2010 Lot# 00327A, 00518A, 00623A, 00831B, 01013A, 01103A, 01214A. 2011 Lot# 10110A, 10301B, 10423A, 10530A, 10602A, 10703A, 10804A, 11004A. 2012 Lot# 120104, 120414.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASystem 1 Base, 220/240V. The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the25 de julio de 2012
- FDAEkoSonic MACH4 Endovascular Device (Catalog #500-55106, 500-55112; 500-55118; 500-55124; 500-55130; 500-55140;25 de julio de 2012
- FDASystem 1 Base, 100/120V. The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the25 de julio de 2012
- FDAOEC 9900 Elite, OEC 9900 Elite MD Motorized C-arm System, OEC 9900 Elite NAV used as an aid to locate25 de julio de 2012
- FDASpacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. It is a patient monitor25 de julio de 2012
- FDAOEC 9800, OEC FluoroTrak 9800 Plus, OEC 9800 Plus, OEC 9800 MD Motorized C-arm System designed to provide25 de julio de 2012
Preguntas sobre este retiro
¿LMA Esophageal Intubation Detector (EID), Catalog Number EID100 The device is used to assist verification of placement of the endotracheal tube or esophageal/tracheal double lumen tube. This device is to be used as an adjunct to assess intubation. Its purpose is not to eliminate clinical judgment fue retirado del mercado?
Sí. FDA publicó el retiro Z-2022-2012 el 25 de julio de 2012. Peligro reportado: The recall was initiated because LMANA has received a complaint that the Esophageal Intubation Detector (EID100) has a potential to create a false positive if used on a patient during confirmation of intubation.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter