Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS, N750-BMKXT-FS, N750-BMKLT-FS, N750-DIS-FS, N750-DISXT-FS. Product Usage: Usage: The Benchmark product lines are intended to automatically stain histological or cytological specimens on microscope slides with specific immunohistochemistry or in situ hybridization reagents for in vitro diagnostic use. The Discovery Systems are research use only instruments with similar functionality.
Publicado el 4 de septiembre de 2013 · Inicio del retiro: 12 de junio de 2008
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Ventana Medical System is initiating the recall of the BenchMark and DISCOVERY series instruments because there is a possibility that the heating element could separate from the underside of the heater cap and cause a non-staining or inconsistent staining event. Additionally, the system may not detect this anomaly and therefore it may not generate an error message for the slide position.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Ventana Medical Systems Inc
- Distribuidor
- Worldwide Distribution - USA Nationwide and the countries of Canada, Europe, Asia, Japan.
- Lote
- BenchMark XT Instruments-S/N: 711869-712465; BenchMark LT instruments-S/N: 610321-610345. Discovery XT Instruments-S/N: 610321-610345.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States