SIGNA Architect

Publicado el 30 de julio de 2025 · Inicio del retiro: 23 de mayo de 2025

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
GE Medical Systems, LLC
Distribuidor
Worldwide distribution - US Nationwide and the country of China.
Lote
GTIN: 00840682147095 00840682122702 00195278023643 00840682123440
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States