HORIBA custom configured fluorescence instrument, modular Fluorolog-QM
Publicado el 10 de julio de 2024 · Inicio del retiro: 28 de marzo de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
HORIBA custom configured fluorescence instrument has a laser safety defect. This custom configuration of the modular Instrument removed the primary light source and the excitation monochromator. A plastic cap was installed to plug the optical port, where a more durable beam stop should have been used. The defect may allow the class 4 laser beam to burn (or transmit) through the plastic cap to exit the protective enclosure.
Qué hacer
FDA Mandated · Letter
Detalles del retiro
- Fabricante
- Horiba Instruments Incorporated
- Distribuidor
- US Nationwide and Worldwide Distribution
- Lote
- Modular Fluorolog-QM
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?