FDAFDA · Alimentos y medicamentos·N.º oficial Z-2167-2018
The VITEK¿ 2 Gram Positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents. Fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Publicado el 20 de junio de 2018 · Inicio del retiro: 23 de marzo de 2018
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- BioMerieux SA
- Distribuidor
- International distribution.
- Lote
- VITEK¿ 2 AST-P606 (ref: 22330): Lot Number 4860117403
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- France
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.