ABX PENTRA Magnesium RTU Reagent, Part No. A11A01646 Product Usage: ABX PENTRA Magnesium RTU reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of magnesium in human serum and plasma based on a photometric test using xylidyl blue. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemnia (abnormally low plasma levels of magnesium) and hypermagnesemnia (abnormally high plasma levels of magnesium).
Publicado el 1 de octubre de 2014 · Inicio del retiro: 11 de agosto de 2014
Resumen
FDA registra el retiro Z-2692-2014 de dispositivos médicos de Clase II, publicado el 1 de octubre de 2014 e iniciado el 11 de agosto de 2014. Empresa responsable: Horiba Instruments, Inc dba Horiba Medical; Estado: Finalizado; Cantidad reportada: 359 units; Lote: All Lots. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Horiba Medical is recalling the ABX PENTRA Magnesium RTU Reagent because the on-board stability claims has been updated. Horiba has updated the ABX PENTRA Magnesium RTU package inserts by removing the 7-day on-board stability claim and replacing it with a 1-day on-board stability claim.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Horiba Instruments, Inc dba Horiba Medical
- Distribuidor
- US Nationwide Distribution.
- Lote
- All Lots
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in conjunction with4 de julio de 2018
- FDAJustright Surgical, JustRight 5mm Reload, REFJR-REL25-2.0, Sterile, EO, Rx Only It is a replacement cartridge1 de octubre de 2014
- FDAMinstrel (with scale) Product Usage: is a mobile passive hoist. It is intended for lifting and transporting1 de octubre de 2014
- FDADevice Name: ABX PENTRA ML Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versions19 de marzo de 2014
- FDAABX PENTRA Multical. ABX PENTRA N Control. ABX PENTRA P Control. The ABX PENTRA 400 and PC200 are benchtop12 de marzo de 2014
- FDAABX PENTRA N Control, Model No. A11A01653 The ABX PENTRA N Control is for use in quality control by monitoring12 de febrero de 2014
Preguntas sobre este retiro
¿ABX PENTRA Magnesium RTU Reagent, Part No. A11A01646 Product Usage: ABX PENTRA Magnesium RTU reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of magnesium in human serum and plasma based on a photometric test using xylidyl blue. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemnia (abnormally low plasma levels of magnesium) and hypermagnesemnia (abnormally high plasma levels of magnesium). fue retirado del mercado?
Sí. FDA publicó el retiro Z-2692-2014 el 1 de octubre de 2014. Peligro reportado: Horiba Medical is recalling the ABX PENTRA Magnesium RTU Reagent because the on-board stability claims has been updated. Horiba has updated the ABX PENTRA Magnesium RTU package inserts by removing the 7-day on-board stability claim and replacing it with a 1-day on-board stability claim.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter