Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2793-2026

SafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix System

Publicado el 29 de julio de 2026 · Inicio del retiro: 13 de mayo de 2026

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of communication between the Patient Interface and Tablet, and prompts requiring system reboot; may result in increased operative time and surgical plan change.

Qué hacer

Voluntary: Firm initiated

Detalles del retiro

Fabricante
Alphatec Spine, Inc.
Distribuidor
US Nationwide distribution in the states of TX, MO, IN.
Lote
UDI-DI: 00190376517506. System Set ID/Patient Interface Serial Number: 2007716/ DG106225, 2005498/ DG106220, 2005490/ DG106230
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta