Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-3262-2024

PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended to provide high flow warmed and humidified respiratory gases for administration to spontaneously breathing infant, child, adolescent and adult patients in hospitals and sub-acute facilities.

Publicado el 2 de octubre de 2024 · Inicio del retiro: 7 de agosto de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Fisher & Paykel Healthcare, Ltd.
Distribuidor
U.S.: AZ, CA, CO, CT, DC, FL, GA, IA, IL, KY, MA, MD, MI, MN, MO, MT, ND, NH, NJ, NV, NY, OH, OR, PA, SD, TX, UT, VA and WA. O.U.S.: Not provided
Lote
Model Number: PT301US UDI-DI code: 09420012466662 Lot/Serial Numbers:
Clasificación de riesgo
Class II
Área (FDA)
device
País
New Zealand

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta