Curaplex Epi Safe Kit, 8600-01102 TRAINING ONLY, Rx Only, Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc. 5000 Tuttle Crossing Blvd. Dublin, OH 43016
Published January 30, 2019 · Recall initiated: November 2, 2018
Summary
FDA records drug Class II recall D-0385-2019, published January 30, 2019 and initiated November 2, 2018. Recalling firm: Bound Tree Medical, LLC; Status: Terminated; Reported quantity: 2,654 kits (6 syringes per kit / 1 IFU per kit); Lot: all lot numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Labeling Incorrect Instructions: This recall has been initiated because the Instructions for Use included in the Epi-Safe Kits for the Epi-Safe Syringe (and/or the Safety Lok Syringe in Model 8600-01120) recommend a midpoint dosage of epinephrine for children between 0 lbs. and 66 lbs. of 0.15mL that has not been approved by FDA for children. Reliance on the current IFU and/or use of the Epi-Safe Syringe as instructed may result in the administration of an inappropriate dosage of epinephrine to young children experiencing anaphylaxis, especially those under 33 lbs. Additionally the Epi Safe syringe does not have the correct markings to allow a health care provider to draw up the correct dosage for children under 33 kg.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Bound Tree Medical, LLC
- Distributor
- Nationwide in the USA
- Lot
- all lot numbers
- Risk classification
- Class II
- Area (FDA)
- drug
- Country
- United States
Related recalls
- FDACuraplex Epi Kit NOT FOR IV USE, #8600-01120, Epi-Safe Kit 8600-01 120 contains:2 Safety-LOK Syringe, TB, 1ccJanuary 30, 2019
- FDAcuraplex Epi Safe Administration and Training Kits, #8600-01100, Contains: 2 Epi Safe Administration KitJanuary 30, 2019
- FDAcuraplex Epi Safe Kit, 8600-01101, contains:" 1ml Vial of Epinephrine " 1 Epi-Safe Syringe " 1 Safety Needle "January 30, 2019
- FDALOPERAMIDE HCL TABLETS USP, 2 mg, ANTI-DIARRHEAL, 24 caplets per carton, OTC, Distributed by Drug Mart-FoodJanuary 30, 2019
- FDAcuraplex Epi Safe Administration and Training Kits # 8600-01100. Kit contains 2 Epi Safe Administration KitJanuary 23, 2019
- FDAU by Kotex® Sleek®, Regular Tampons, 3 CountJanuary 23, 2019
Questions about this recall
Was Curaplex Epi Safe Kit, 8600-01102 TRAINING ONLY, Rx Only, Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc. 5000 Tuttle Crossing Blvd. Dublin, OH 43016 recalled?
Yes. FDA published recall D-0385-2019 on January 30, 2019. Reported hazard: Labeling Incorrect Instructions: This recall has been initiated because the Instructions for Use included in the Epi-Safe Kits for the Epi-Safe Syringe (and/or the Safety Lok Syringe in Model 8600-01120) recommend a midpoint dosage of epinephrine for children between 0 lbs. and 66 lbs. of 0.15mL that has not been approved by FDA for children. Reliance on the current IFU and/or use of the Epi-Safe Syringe as instructed may result in the administration of an inappropriate dosage of epinephrine to young children experiencing anaphylaxis, especially those under 33 lbs. Additionally the Epi Safe syringe does not have the correct markings to allow a health care provider to draw up the correct dosage for children under 33 kg.
What should I do if I have this product?
Voluntary: Firm initiated · Letter