Human Alpha-1-Antitrypsin Kit for use on the SPAPlus analyser, Product #NK034.S. The firm name on the label is The Binding Site Group Ltd., Birmingham, UK. Product is designed for the quantitative in-vitro determination of Alpha-1-antitrypsin in human serum using the SPAPLUS turbidimetric analyser. The measurement of Alpha-1-antitrypsin aids in the diagnosis of several conditions including adult cirrhosis of the liver. In addition, Alpha-1-antitrypsin deficiency has been associated with pulmonary emphysema. This test should be used in conjunction with other laboratory and clinical findings.
Published October 11, 2017 · Recall initiated: August 23, 2017
Summary
FDA records medical device Class II recall Z-0002-2018, published October 11, 2017 and initiated August 23, 2017. Recalling firm: The Binding Site Group, Ltd.; Status: Terminated; Reported quantity: 361 kits; Lot: Lot #400900, Exp. 12/31/2017, UDI #05051700016421. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The high control for the kit is recovering low and may be out of range for assigned values.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- The Binding Site Group, Ltd.
- Distributor
- Worldwide Distribution - US Distribution to the state of CA., and to the countries of : United Kingdom, Germany, France, Czech Republic, Belgium, Panama, Bulgaria, Iran, Argentina, Turkey Romania, Colombia, and Saudi Arabia.
- Lot
- Lot #400900, Exp. 12/31/2017, UDI #05051700016421
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDABrand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20June 17, 2026
- FDAThe Optilite Clinical Chemistry Analyzer. IVD. Model Number: IE700. UDI-05051700017176 - Product Usage: TheMay 27, 2020
- FDAOPTILITE IgM Kit, REF: NK012.OPT.A, IVD, UDI: 05051700017565February 26, 2020
- FDABinding Site Human IgM Kit (for SPAPLUS), REF: NK012.S The device consists of reaction buffer, sheep anti-IgMDecember 19, 2018
- FDACarl Zeiss INTRABEAM PRS 500, product #304534-0000-500. The firm name on the label is Carl Zeiss Meditec AG,October 11, 2017
- FDAElectric Wheelchair bases: a) C300 b) C350 c) C400 d) C500October 11, 2017
Questions about this recall
Was Human Alpha-1-Antitrypsin Kit for use on the SPAPlus analyser, Product #NK034.S. The firm name on the label is The Binding Site Group Ltd., Birmingham, UK. Product is designed for the quantitative in-vitro determination of Alpha-1-antitrypsin in human serum using the SPAPLUS turbidimetric analyser. The measurement of Alpha-1-antitrypsin aids in the diagnosis of several conditions including adult cirrhosis of the liver. In addition, Alpha-1-antitrypsin deficiency has been associated with pulmonary emphysema. This test should be used in conjunction with other laboratory and clinical findings. recalled?
Yes. FDA published recall Z-0002-2018 on October 11, 2017. Reported hazard: The high control for the kit is recovering low and may be out of range for assigned values.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail