Back to search
FDAFDA · Food and drugs·Recall number Z-0008-2014

Draeger Fabius GS Premium, Fabius OS, Fabius Tiro, Fabius Tiro D-M Product Usage: used to administer anesthesia and ventilation to patients during surgical procedures.

Published October 16, 2013 · Recall initiated: August 14, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During final testing of the Draeger Fabius GS Premium, Fabius OS, and Fabius Tiro anesthesia machines, some units did not pass the high voltage test and the minimum clearance between an electrical component and the unit housing was not maintained for some from a specific batch.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Draeger Medical, Inc.
Distributor
Worldwide distribution: US (nationwide) and countries of: Argentina, Australia, Austria, Azerbaijan, Bangladesh, Brazil, Bulgaria, Canada, Chile, China, Columbia, Ecuador, Egypt, France, Germany, Greece, Hungary, India, Indonesia, Israel, Italy, Ivory Coast, Japan, Kazakhstan, Latvia, Libya, Lithuania Malaysia, Morocco, Myanmar, Nepal, Philippines, Poland, Rumania, Russian Fed., Saudi Arabia, Senegal, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Uruguay, Vietnam, White Russia Turkmenistan.
Lot
Part Nos. 8607000 (GS Permium) 8606000 (Fabius OS and Fabius Tiro) 8608400 (Fabius Tiro D-M)ASEC-0055 with lot numbers: ASEC-0186, ASED-0006, ASED-0074, ASED-0291, ASEC-0056, ASEC-0205, ASED-0007, SED-0075 ASED-0292, ASEC-0061, ASEC-0206, ASED-0020, ASED-0079, ASED-0293, ASEC-0062, ASEC-0207, ASED-0021, ASED-0139, ASED-0294, ASEC-0070, ASEC-0208, ASED-0023, ASED-0160, ASED-0295, ASEC-0071, ASEC-0209, ASED-0024, ASED-0161, ASED-0296, ASEC-0112, ASEC-0210, ASED-0025, ASED-0162, ASEE-0021, ASEC-0124, SEC-0211, ASED-0031, ASED-0163, ASEE-0022, ASEC-0125, ASEC-0212, ASED-0032, ASED-0164, ASEE-0023, ASEC-0126, ASEC-0213, ASED-0033, ASED-0165, ASEE-0041, ASEC-0169, ASEC-0214, ASED-0034, ASED-0166, ASEE-0042, ASEC-0170, ASEC-0236, ASED-0035, ASED-0167, ASEE-0058, ASEC-0171, ASEC-0248, ASED-0036, ASED-0168, ASEE-0060, ASEC-0172, ASEC-0249, ASED-0037, ASED-0169, ASEE-0061, ASEC-0173, ASEC-0250, ASED-0038, ASED-0203, ASEE-0065, ASEC-0174, ASED-0001, ASED-0044, ASED-0253, ASEE-0066, ASEC-0175, ASED-0002, ASED-0045, ASED-0286, ASEE-0083, ASEC-0176, ASED-0003, ASED-0046, ASED-0287, ASEE-0103, ASEC-0184, ASED-0004, ASED-0047, ASED-0288, ASEE-0136, ASEC-0185, ASED-0005, ASED-0051, ASED-0290, ASEC-0054, ASEC-0194, ASED-0247, ASED-0263, ASEE-0072, ASEC-0167, ASEC-0195, ASED-0248, ASED-0264, ASEE-0089, ASEC-0178, ASEC-0196, ASED-0254, ASED-0265, ASEE-0090, ASEC-0179, ASEC-0197, ASED-0255, ASED-0266, ASEE-0131, ASEC-0180, ASEC-0227, ASED-0256, ASED-0282, ASEE-0132, ASEC-0181, ASEC-0228, ASED-0257, ASEE-0003, ASEE-0133, ASEC-0182, ASED-0040, ASED-0260, ASEE-0017, ASEE-0134, ASEC-0183, ASED-0086, ASED-0261, ASEE-0054, ASEC-0193, ASED-0245, ASED-0262, ASEE-0071, ASED-0141, ASED-0143, ASED-0145, ASED-0147, ASED-0149, ASED-0142, ASED-0144, ASED-0146, ASED-0148, ASEB-0219
Risk classification
Class I
Area (FDA)
device
Country
United States