Back to search
FDAFDA · Food and drugs·Recall number Z-0013-2020

Vista Dental Products Sodium Hypochlorite - 3%, 10 - 12cc Pre-filled Syringes, REF 502315

Published October 9, 2019 · Recall initiated: June 27, 2019

Summary

FDA records medical device Class II recall Z-0013-2020, published October 9, 2019 and initiated June 27, 2019. Recalling firm: Inter-Med Llc; Status: Terminated; Reported quantity: 582 units; Lot: Lot/Work Order Numbers: 2018-1349 2018-1587 2018-1853 20182107, UDI (01)10818207020427 20182303, UDI (01)10818207020427 20182584, UDI (01)10818207020427 20182666, UDI (01)10818207020427 20190199, UDI (01)10818207020427 20190433, UDI (01)10818207020427. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Inter-Med Llc
Distributor
US, South Africa, Switzerland, Poland, Italy, Singapore, Australia
Lot
Lot/Work Order Numbers: 2018-1349 2018-1587 2018-1853 20182107, UDI (01)10818207020427 20182303, UDI (01)10818207020427 20182584, UDI (01)10818207020427 20182666, UDI (01)10818207020427 20190199, UDI (01)10818207020427 20190433, UDI (01)10818207020427
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Vista Dental Products Sodium Hypochlorite - 3%, 10 - 12cc Pre-filled Syringes, REF 502315 recalled?

Yes. FDA published recall Z-0013-2020 on October 9, 2019. Reported hazard: The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.

What should I do if I have this product?

Voluntary: Firm initiated · Letter