Vista Dental Products Sodium Hypochlorite - 3%, 10 - 12cc Pre-filled Syringes, REF 502315
Published October 9, 2019 · Recall initiated: June 27, 2019
Summary
FDA records medical device Class II recall Z-0013-2020, published October 9, 2019 and initiated June 27, 2019. Recalling firm: Inter-Med Llc; Status: Terminated; Reported quantity: 582 units; Lot: Lot/Work Order Numbers: 2018-1349 2018-1587 2018-1853 20182107, UDI (01)10818207020427 20182303, UDI (01)10818207020427 20182584, UDI (01)10818207020427 20182666, UDI (01)10818207020427 20190199, UDI (01)10818207020427 20190433, UDI (01)10818207020427. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Inter-Med Llc
- Distributor
- US, South Africa, Switzerland, Poland, Italy, Singapore, Australia
- Lot
- Lot/Work Order Numbers: 2018-1349 2018-1587 2018-1853 20182107, UDI (01)10818207020427 20182303, UDI (01)10818207020427 20182584, UDI (01)10818207020427 20182666, UDI (01)10818207020427 20190199, UDI (01)10818207020427 20190433, UDI (01)10818207020427
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAdvance Aluminum Chloride Gel Clear Hemostatic Gel Kit, Catalog number 504600-28June 17, 2026
- FDAVista Dental Products Sodium Hypochlorite - 6%, QTY: 12, 3mL Pre-filled Syringes, REF 502350October 9, 2019
- FDAExactaMix, Model: H938905, EMPTY EVA DUAL CHAMBER CONTAINER With Screw Connectors, 250 ml/ 1250 ml, Baxter -October 9, 2019
- FDASECURE EVA - DUAL CHAMBER CONTAINER with Manifold, 1500mL, REF 66616 (Canadian distribution) - Product Usage:October 9, 2019
- FDAPaltop Advanced Plus Dental Implant Dia 6.0mm L10mm Catalog Number: 20-70024POctober 9, 2019
- FDASECURE EVA - DUAL CHAMBER CONTAINER With Screw Connectors, 4000mL, (legless) REF 66640 - Product Usage: ForOctober 9, 2019
Questions about this recall
Was Vista Dental Products Sodium Hypochlorite - 3%, 10 - 12cc Pre-filled Syringes, REF 502315 recalled?
Yes. FDA published recall Z-0013-2020 on October 9, 2019. Reported hazard: The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.
What should I do if I have this product?
Voluntary: Firm initiated · Letter