FDAFDA · Food and drugs·Recall number Z-0014-2025
CADD Medication Cassette Reservoir, 50mL, Product Code 21-7001-24
Published October 30, 2024 · Recall initiated: June 26, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Smiths Medical ASD Inc.
- Distributor
- Worldwide distribution.
- Lot
- UDI/DI 10610586022111, Lot Numbers: 4040237, 4013368, 3983342, 4046817, 405240, 4060903
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States