Back to search
FDAFDA · Food and drugs·Recall number Z-0022-2025

FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer

Published October 16, 2024 · Recall initiated: August 26, 2024

Summary

FDA records medical device Class II recall Z-0022-2025, published October 16, 2024 and initiated August 26, 2024. Recalling firm: Qualigen Inc; Status: Ongoing; Reported quantity: 60; Lot: UDI-DI: 00816467020099, Lot: 2405017-1, Expiration: 2025-05-15. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Qualigen Inc
Distributor
US: TX, MS, NY, OH, MT, WA, PA, MN, SC, GA, NC, NV, WY, AK, SD, KS, NE, VT, AL, MA, OK, UT, WI, IA, WV. OUS: Germany, Switzerland
Lot
UDI-DI: 00816467020099, Lot: 2405017-1, Expiration: 2025-05-15
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer recalled?

Yes. FDA published recall Z-0022-2025 on October 16, 2024. Reported hazard: Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.

What should I do if I have this product?

Voluntary: Firm initiated · Letter