Infinity 5 Implantable Pulse Generator and Infinity 7 Implantable Pulse Generator Models 6660, 6661, 6662, 6663
Published October 25, 2017 · Recall initiated: September 12, 2017
Summary
FDA records medical device Class II recall Z-0023-2018, published October 25, 2017 and initiated September 12, 2017. Recalling firm: St. Jude Medical, Inc.; Status: Terminated; Reported quantity: 3,119 devices; Lot: Serials implanted prior to September 12, 2017 are considered part of this advisory; Serial Numbers - AAD193.1 AAD194.1 AAD213.1 AAD255.1 AAD257.1 AAD261.1 AAD276.1 AAD279.1 AAD281.1 AAD298.1 AAD299.1 AAD302.1 AAD309.1 AAD328.1 AAD372.1 AAD376.1 AAD380.1 AAN940.1 AAN946.1 AAN951.1 AAN977.1 AAN997.1 AAP006.1 AAP007.1 AAP030.1 AAP035.1 AAP108.1 AAP132.1 AAP140.1 AAP149.1 AAP193.1 AAP217.1 AAP266.1 AAP273.1 AAP279.1 AAP293.1 AAP314.1 AAP328.1 AAP333.1 AAP362.1 AAP363.1 AAP375.1 AAP376.1 AAP385.1 AAP387.1 AAP409.1 AAP410.1 AAT509.1 AAT540.1 AAT571.1 AAT572.1 AAT577.1 AAT579.1 AAT582.1 AAT626.1 AAT634.1 AAT652.1 AAT657.1 AAT660.1 AAT722.1 AAT754.1 AAT772.1 AAT774.1 AAT776.1 AAT783.1 AAT792.1 AAT795.1 AAT797.1 AAT804.1 AAT805.1 AAT830.1 AAT841.1 AAT845.1 AAT849.1 AAT870.1 AAT873.1 AAT885.1 AAT890.1 AAT891.1 AAT899.1 AAT902.1 AAT910.1 AAT976.1 AAT987.1 AAT989.1 AAT990.1 AAT994.1 AAU003.1 AAU009.1 AAU022.1 AAU029.1 AAU031.1 AAU046.1 AAU047.1 AAU056.1 AAU068.1 AAU071.1 AAU084.1 AAU094.1 AAU095.1 AAU096.1 AAU148.1 ABA946.1 ABA964.1 ABA995.1 ABB011.1 ABB022.1 ABB051.1 ABB128.1 ABJ141.1 ABJ142.1 ABJ148.1 ABJ155.1 ABJ156.1 ABJ165.1 ABJ175.1 ABJ177.1 ABJ186.1 ABJ196.1 ABJ197.1 ABJ234.1 ABJ242.1 ABJ243.1 ABJ255.1 ABJ282.1 ABJ284.1 ABJ285.1 ABJ295.1 ABJ311.1 ABJ313.1 ABJ315.1 ABJ321.1 ABJ328.1 ABJ349.1 ABJ386.1 ABJ388.1 ABJ390.1 ABJ392.1 ABJ411.1 ABJ444.1 ABJ445.1 ABJ448.1 ABJ449.1 ABJ458.1 ABJ472.1 ABJ485.1 ABJ501.1 ABJ503.1 ABJ504.1 ABJ505.1 ABJ506.1 ABJ526.1 ABJ556.1 ABJ614.1 ABM287.1 ABM308.1 ABM310.1 ABM312.1 ABM313.1 ABM329.1 ABM334.1 ABM381.1 ABM408.1 ABM459.1 ABM471.1 ABM472.1 ABM540.1 ABM541.1 ABM581.1 ABM589.1 ABM592.1 ABM593.1 ABM620.1 ABM630.1 ABM651.1 ABT714.1 ABT728.1 ABT739.1 ABT740.1 ABT745.1 ABT785.1 ABT798.1 ABT800.1 ABT811.1 ABT829.1 ABT832.1 ABT839.1 ABT841.1 ABT844.1 ABT883.1 ABT892.1 ABT900.1 ABT948.1 ABT956.1 ABT967.1 ABT975.1 ABT981.1 ABT984.1 ABT990.1 ABT991.1 ABU014.1 ABU018.1 ABU024.1 ABU036.1 ABU058.1 ABU061.1 ABU079.1 ABU099.1 ABU107.1 ABU119.1 ABU146.1 ABU164.1 ABU183.1 ACD615.1 ACD617.1 ACD652.1 ACD655.1 ACD656.1 ACD682.1 ACD686.1 ACD721.1 ACD768.1 ACD772.1 ACD773.1 ACD797.1 ACD800.1 ACD802.1 ACD812.1 ACD849.1 ACE661.1 ACE669.1 ACE689.1 ACE699.1 ACE740.1 ACE771.1 ACE772.1 ACE780.1 ACE786.1 ACE838.1 ACE843.1 ACE860.1 ACE868.1 ACE875.1 ACE877.1 ACE892.1 ACE900.1 ACE912.1 ACE917.1 ACE922.1 ACE936.1 ACE939.1 ACE942.1 ACE947.1 ACE994.1 ACF006.1 ACF039.1 ACF048.1 ACF065.1 ACK135.1 ACK136.1 ACK140.1 ACK145.1 ACK159.1 ACK161.1 ACK165.1 ACK200.1 ACK224.1 ACK228.1 ACK231.1 ACK241.1 ACK253.1 ACK269.1 ACK286.1 ACK289.1 ACK318.1 ACK325.1 ACK379.1 ACK385.1 ACK424.1 ACK428.1 ACK429.1 ACK431.1 ACK448.1 ACK457.1 ACK500.1 ACK511.1 ACK514.1 ACK525.1 ACK526.1 ACK549.1 ACK554.1 ACK559.1 ACK581.1 ACK583.1 ACK592.1 ACK594.1 ACK602.1 ACK615.1 ACK622.1 ACK629.1 ACK632.1 ACK634.1 ACK637.1 ACK674.1 ACK677.1 ACK699.1 ACK702.1 ACK705.1 ACK706.1 ACK709.1 ACK719.1 ACK739.1 ACK742.1 ACK744.1 ACK752.1 ACK757.1 ACK759.1 ACK760.1 ACK762.1 ACK775.1 ACK776.1 ACK778.1 ACK783.1 ACK784.1 ACK786.1 ACK803.1 ACK805.1 ACK806.1 ACK828.1 ACK837.1 ACK841.1 ACK844.1 ACK846.1 ACK855.1 ACK872.1 ACP669.1 ACP670.1 ACP675.1 ACP676.1 ACP680.1 ACP684.1 ACP685.1 ACP686.1 ACP687.1 ACP690.1 ACP695.1 ACP700.1 ACP737.1 ACP756.1 ACP767.1 ACP769.1 ACP770.1 ACP775.1 ACP777.1 ACP780.1 ACP783.1 ACP797.1 ACP804.1 ACP820.1 ACP824.1 ACP827.1 ACP829.1 ACP833.1 ACP835.1 ACP836.1 ACP862.1 ACP866.1 ACP869.1 ACP872.1 ACP874.1 ACP881.1 ACP883.1 ACP885.1 ACP890.1 ACP893.1 ACP909.1 ACP915.1 ACT538.1 ACT542.1 ACT547.1 ACT548.1 ACT555.1 ACT556.1 ACT558.1 ACT562.1 ACT563.1 ACT565.1 ACT580.1 ACT582.1 ACT589. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- St. Jude Medical, Inc.
- Distributor
- Worldwide Distribution - US (nationwide) (foreign consignee list will be provide by the firm at a later time)
- Lot
- Serials implanted prior to September 12, 2017 are considered part of this advisory; Serial Numbers - AAD193.1 AAD194.1 AAD213.1 AAD255.1 AAD257.1 AAD261.1 AAD276.1 AAD279.1 AAD281.1 AAD298.1 AAD299.1 AAD302.1 AAD309.1 AAD328.1 AAD372.1 AAD376.1 AAD380.1 AAN940.1 AAN946.1 AAN951.1 AAN977.1 AAN997.1 AAP006.1 AAP007.1 AAP030.1 AAP035.1 AAP108.1 AAP132.1 AAP140.1 AAP149.1 AAP193.1 AAP217.1 AAP266.1 AAP273.1 AAP279.1 AAP293.1 AAP314.1 AAP328.1 AAP333.1 AAP362.1 AAP363.1 AAP375.1 AAP376.1 AAP385.1 AAP387.1 AAP409.1 AAP410.1 AAT509.1 AAT540.1 AAT571.1 AAT572.1 AAT577.1 AAT579.1 AAT582.1 AAT626.1 AAT634.1 AAT652.1 AAT657.1 AAT660.1 AAT722.1 AAT754.1 AAT772.1 AAT774.1 AAT776.1 AAT783.1 AAT792.1 AAT795.1 AAT797.1 AAT804.1 AAT805.1 AAT830.1 AAT841.1 AAT845.1 AAT849.1 AAT870.1 AAT873.1 AAT885.1 AAT890.1 AAT891.1 AAT899.1 AAT902.1 AAT910.1 AAT976.1 AAT987.1 AAT989.1 AAT990.1 AAT994.1 AAU003.1 AAU009.1 AAU022.1 AAU029.1 AAU031.1 AAU046.1 AAU047.1 AAU056.1 AAU068.1 AAU071.1 AAU084.1 AAU094.1 AAU095.1 AAU096.1 AAU148.1 ABA946.1 ABA964.1 ABA995.1 ABB011.1 ABB022.1 ABB051.1 ABB128.1 ABJ141.1 ABJ142.1 ABJ148.1 ABJ155.1 ABJ156.1 ABJ165.1 ABJ175.1 ABJ177.1 ABJ186.1 ABJ196.1 ABJ197.1 ABJ234.1 ABJ242.1 ABJ243.1 ABJ255.1 ABJ282.1 ABJ284.1 ABJ285.1 ABJ295.1 ABJ311.1 ABJ313.1 ABJ315.1 ABJ321.1 ABJ328.1 ABJ349.1 ABJ386.1 ABJ388.1 ABJ390.1 ABJ392.1 ABJ411.1 ABJ444.1 ABJ445.1 ABJ448.1 ABJ449.1 ABJ458.1 ABJ472.1 ABJ485.1 ABJ501.1 ABJ503.1 ABJ504.1 ABJ505.1 ABJ506.1 ABJ526.1 ABJ556.1 ABJ614.1 ABM287.1 ABM308.1 ABM310.1 ABM312.1 ABM313.1 ABM329.1 ABM334.1 ABM381.1 ABM408.1 ABM459.1 ABM471.1 ABM472.1 ABM540.1 ABM541.1 ABM581.1 ABM589.1 ABM592.1 ABM593.1 ABM620.1 ABM630.1 ABM651.1 ABT714.1 ABT728.1 ABT739.1 ABT740.1 ABT745.1 ABT785.1 ABT798.1 ABT800.1 ABT811.1 ABT829.1 ABT832.1 ABT839.1 ABT841.1 ABT844.1 ABT883.1 ABT892.1 ABT900.1 ABT948.1 ABT956.1 ABT967.1 ABT975.1 ABT981.1 ABT984.1 ABT990.1 ABT991.1 ABU014.1 ABU018.1 ABU024.1 ABU036.1 ABU058.1 ABU061.1 ABU079.1 ABU099.1 ABU107.1 ABU119.1 ABU146.1 ABU164.1 ABU183.1 ACD615.1 ACD617.1 ACD652.1 ACD655.1 ACD656.1 ACD682.1 ACD686.1 ACD721.1 ACD768.1 ACD772.1 ACD773.1 ACD797.1 ACD800.1 ACD802.1 ACD812.1 ACD849.1 ACE661.1 ACE669.1 ACE689.1 ACE699.1 ACE740.1 ACE771.1 ACE772.1 ACE780.1 ACE786.1 ACE838.1 ACE843.1 ACE860.1 ACE868.1 ACE875.1 ACE877.1 ACE892.1 ACE900.1 ACE912.1 ACE917.1 ACE922.1 ACE936.1 ACE939.1 ACE942.1 ACE947.1 ACE994.1 ACF006.1 ACF039.1 ACF048.1 ACF065.1 ACK135.1 ACK136.1 ACK140.1 ACK145.1 ACK159.1 ACK161.1 ACK165.1 ACK200.1 ACK224.1 ACK228.1 ACK231.1 ACK241.1 ACK253.1 ACK269.1 ACK286.1 ACK289.1 ACK318.1 ACK325.1 ACK379.1 ACK385.1 ACK424.1 ACK428.1 ACK429.1 ACK431.1 ACK448.1 ACK457.1 ACK500.1 ACK511.1 ACK514.1 ACK525.1 ACK526.1 ACK549.1 ACK554.1 ACK559.1 ACK581.1 ACK583.1 ACK592.1 ACK594.1 ACK602.1 ACK615.1 ACK622.1 ACK629.1 ACK632.1 ACK634.1 ACK637.1 ACK674.1 ACK677.1 ACK699.1 ACK702.1 ACK705.1 ACK706.1 ACK709.1 ACK719.1 ACK739.1 ACK742.1 ACK744.1 ACK752.1 ACK757.1 ACK759.1 ACK760.1 ACK762.1 ACK775.1 ACK776.1 ACK778.1 ACK783.1 ACK784.1 ACK786.1 ACK803.1 ACK805.1 ACK806.1 ACK828.1 ACK837.1 ACK841.1 ACK844.1 ACK846.1 ACK855.1 ACK872.1 ACP669.1 ACP670.1 ACP675.1 ACP676.1 ACP680.1 ACP684.1 ACP685.1 ACP686.1 ACP687.1 ACP690.1 ACP695.1 ACP700.1 ACP737.1 ACP756.1 ACP767.1 ACP769.1 ACP770.1 ACP775.1 ACP777.1 ACP780.1 ACP783.1 ACP797.1 ACP804.1 ACP820.1 ACP824.1 ACP827.1 ACP829.1 ACP833.1 ACP835.1 ACP836.1 ACP862.1 ACP866.1 ACP869.1 ACP872.1 ACP874.1 ACP881.1 ACP883.1 ACP885.1 ACP890.1 ACP893.1 ACP909.1 ACP915.1 ACT538.1 ACT542.1 ACT547.1 ACT548.1 ACT555.1 ACT556.1 ACT558.1 ACT562.1 ACT563.1 ACT565.1 ACT580.1 ACT582.1 ACT589.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips Healthcare Brilliance iCT X-Ray, Tomography, Computed These computed tomography x-ray systems areOctober 25, 2017
- FDAAlaris Pump Module model 8100 manufactured between November 2011 and March 2012; Alaris Pump Module servicedOctober 25, 2017
- FDAProclaim 5 Elite, implantable Pulse Generator, REF 3660, Rx only, STERILE EO Product Usage: The ProclaimJuly 19, 2017
- FDAProclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO Product Usage: The ProclaimJuly 19, 2017
- FDAInfinity 5, Implantable Pulse Generator, REF 6661, Rx only, STERILE EO Product Usage: The Infinity IPG is anJuly 19, 2017
- FDAInfinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO Product Usage: The Infinity IPG is anJuly 19, 2017
Questions about this recall
Was Infinity 5 Implantable Pulse Generator and Infinity 7 Implantable Pulse Generator Models 6660, 6661, 6662, 6663 recalled?
Yes. FDA published recall Z-0023-2018 on October 25, 2017. Reported hazard: Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations.
What should I do if I have this product?
Voluntary: Firm initiated · Letter