Zimmer Universal Locking System, Titanium 3.5mm. Zimmer Implant and Instrument Cases are reusable, non-implantable devices designed to facilitate organization, identification, storage, transportation, cleaning, and sterilization processing of reusable instruments and certain unused implantable devices.

Published October 14, 2015 · Recall initiated: August 11, 2015

Summary

FDA records medical device Class II recall Z-0027-2016, published October 14, 2015 and initiated August 11, 2015. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 18 units; Lot: Item No. 00-4836-080-00; Lot 56555128. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

It was found through review of open Complaints and stock investigations that a single lot of Zimmer Universal Locking System 3.5 mm Stainless Steel Base and Trays has incorrect silk screening as applied by the supplier. The T-plate catalog numbers noted in the tray are listed as the 00-4946-xxx-xx series (titanium plates); however, the correct numbers should be 00-4936-xxx-xx (stainless steel plat

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Distributed in NC, AL, KS, TX, UT, CA, and SINGAPORE.
Lot
Item No. 00-4836-080-00; Lot 56555128
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Zimmer Universal Locking System, Titanium 3.5mm. Zimmer Implant and Instrument Cases are reusable, non-implantable devices designed to facilitate organization, identification, storage, transportation, cleaning, and sterilization processing of reusable instruments and certain unused implantable devices. recalled?

Yes. FDA published recall Z-0027-2016 on October 14, 2015. Reported hazard: It was found through review of open Complaints and stock investigations that a single lot of Zimmer Universal Locking System 3.5 mm Stainless Steel Base and Trays has incorrect silk screening as applied by the supplier. The T-plate catalog numbers noted in the tray are listed as the 00-4946-xxx-xx series (titanium plates); however, the correct numbers should be 00-4936-xxx-xx (stainless steel plat

What should I do if I have this product?

Voluntary: Firm initiated · Letter