Zimmer Universal Locking System, Titanium 3.5mm. Zimmer Implant and Instrument Cases are reusable, non-implantable devices designed to facilitate organization, identification, storage, transportation, cleaning, and sterilization processing of reusable instruments and certain unused implantable devices.
Published October 14, 2015 · Recall initiated: August 11, 2015
Summary
FDA records medical device Class II recall Z-0027-2016, published October 14, 2015 and initiated August 11, 2015. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 18 units; Lot: Item No. 00-4836-080-00; Lot 56555128. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was found through review of open Complaints and stock investigations that a single lot of Zimmer Universal Locking System 3.5 mm Stainless Steel Base and Trays has incorrect silk screening as applied by the supplier. The T-plate catalog numbers noted in the tray are listed as the 00-4946-xxx-xx series (titanium plates); however, the correct numbers should be 00-4936-xxx-xx (stainless steel plat
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Distributed in NC, AL, KS, TX, UT, CA, and SINGAPORE.
- Lot
- Item No. 00-4836-080-00; Lot 56555128
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDACentricity PACS IW by GE Healthcare Dynamic Imaging Solutions is a device that receives medical imagesOctober 14, 2015
- FDAregard Item Number: 800411, Sterile, NU0480 - Neuro Pk - surgical kit containing cover light handle soft, RX.October 14, 2015
Questions about this recall
Was Zimmer Universal Locking System, Titanium 3.5mm. Zimmer Implant and Instrument Cases are reusable, non-implantable devices designed to facilitate organization, identification, storage, transportation, cleaning, and sterilization processing of reusable instruments and certain unused implantable devices. recalled?
Yes. FDA published recall Z-0027-2016 on October 14, 2015. Reported hazard: It was found through review of open Complaints and stock investigations that a single lot of Zimmer Universal Locking System 3.5 mm Stainless Steel Base and Trays has incorrect silk screening as applied by the supplier. The T-plate catalog numbers noted in the tray are listed as the 00-4946-xxx-xx series (titanium plates); however, the correct numbers should be 00-4936-xxx-xx (stainless steel plat
What should I do if I have this product?
Voluntary: Firm initiated · Letter