FDAFDA · Food and drugs·Recall number Z-0029-2014
Model BP7507 OPTI LION E-Plus Cassettes 25 Cassettes with Sample Probes, OPTIMedical Cassettes are consumable used in the OPTI LION Electrolyte Analyzer intended to be used for the measurement of sodium, potassium, chloride, ionized calcium and pH in samples of whole blood, serum, and plasma.
Published October 30, 2013 · Recall initiated: August 9, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
OPTI LION E-Plus Cassettes were labeled with an incorrect barcode. The incorrect barcode could perfect the performance of one or more parameters as the product ages. Also a Customer complaint from a distributor reported that a customer was getting QC failures.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- OPTI Medical Systems, Inc
- Distributor
- Worldwide distribution: US (nationwide) including states of: CA, MA, MN, and VA; and countries of: China, Guatemala, India, Iraq, South Korea, and Untied Arab Emirates.
- Lot
- Lots 252810, 315815, 319810, 319815, 321817, 367812, 370812
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States