Model BP7507 OPTI LION E-Plus Cassettes 25 Cassettes with Sample Probes, OPTIMedical Cassettes are consumable used in the OPTI LION Electrolyte Analyzer intended to be used for the measurement of sodium, potassium, chloride, ionized calcium and pH in samples of whole blood, serum, and plasma.
Published October 30, 2013 · Recall initiated: August 9, 2013
Summary
FDA records medical device Class II recall Z-0029-2014, published October 30, 2013 and initiated August 9, 2013. Recalling firm: OPTI Medical Systems, Inc; Status: Terminated; Reported quantity: 361 boxes of 25 cassettes each; Lot: Lots 252810, 315815, 319810, 319815, 321817, 367812, 370812. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
OPTI LION E-Plus Cassettes were labeled with an incorrect barcode. The incorrect barcode could perfect the performance of one or more parameters as the product ages. Also a Customer complaint from a distributor reported that a customer was getting QC failures.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- OPTI Medical Systems, Inc
- Distributor
- Worldwide distribution: US (nationwide) including states of: CA, MA, MN, and VA; and countries of: China, Guatemala, India, Iraq, South Korea, and Untied Arab Emirates.
- Lot
- Lots 252810, 315815, 319810, 319815, 321817, 367812, 370812
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABBL (tm) MGIT (tm) Mycobacteria Growth Indicator Tube, part # 245122, packaged in 100 ct cartons and labeledOctober 30, 2013
- FDAPerkin Elmer 226 Sample Collection Device, Serial numbers: 06132903 - 06198900. The PerkinElmer 226 SampleOctober 30, 2013
- FDAFrozen Sliced Carrots, packaged in 2 lbs. bag, sold under Simplot Classic brand. The UPC is 0 71179 1846 3 1.October 30, 2013
- FDAUNITRAC Pneumatic Retraction Arm. A support arm used to hold surgical instruments during surgical procedures.October 30, 2013
- FDAGreen Leaf Lettuce is sold in bulk cases under "Garre Farms" or "Rossi Farms" brand. There are 24 heads ofOctober 30, 2013
- FDABP7561 OPTI CCA-TS B-Lac Cassette (25 per box) Cassettes are consumables used in the OPTI CCA-TS AnalyzerOctober 3, 2012
Questions about this recall
Was Model BP7507 OPTI LION E-Plus Cassettes 25 Cassettes with Sample Probes, OPTIMedical Cassettes are consumable used in the OPTI LION Electrolyte Analyzer intended to be used for the measurement of sodium, potassium, chloride, ionized calcium and pH in samples of whole blood, serum, and plasma. recalled?
Yes. FDA published recall Z-0029-2014 on October 30, 2013. Reported hazard: OPTI LION E-Plus Cassettes were labeled with an incorrect barcode. The incorrect barcode could perfect the performance of one or more parameters as the product ages. Also a Customer complaint from a distributor reported that a customer was getting QC failures.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit