FDAFDA · Food and drugs·Recall number Z-0040-2014

Xoran MiniCAT00202

Published November 6, 2013 · Recall initiated: July 27, 2010

Summary

FDA records medical device Class II recall Z-0040-2014, published November 6, 2013 and initiated July 27, 2010. Recalling firm: Xoran Technologies, Inc.; Status: Terminated; Reported quantity: 320; Lot: minicat. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xorans specifications.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Xoran Technologies, Inc.
Distributor
Worldwide Distribution - Nationwide
Lot
minicat
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Xoran MiniCAT00202 recalled?

Yes. FDA published recall Z-0040-2014 on November 6, 2013. Reported hazard: The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xorans specifications.

What should I do if I have this product?

Voluntary: Firm initiated · Letter