Citrine Triple Quad Mass Spectrometer, Part Number 5063684

Published October 11, 2023 · Recall initiated: August 30, 2023

Summary

FDA records medical device Class II recall Z-0040-2024, published October 11, 2023 and initiated August 30, 2023. Recalling firm: AB Sciex, LLC; Status: Ongoing; Reported quantity: 14; Lot: UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
AB Sciex, LLC
Distributor
US Nationwide.
Lot
UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Citrine Triple Quad Mass Spectrometer, Part Number 5063684 recalled?

Yes. FDA published recall Z-0040-2024 on October 11, 2023. Reported hazard: In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

What should I do if I have this product?

Voluntary: Firm initiated · Letter