Citrine Triple Quad Mass Spectrometer, Part Number 5063684
Published October 11, 2023 · Recall initiated: August 30, 2023
Summary
FDA records medical device Class II recall Z-0040-2024, published October 11, 2023 and initiated August 30, 2023. Recalling firm: AB Sciex, LLC; Status: Ongoing; Reported quantity: 14; Lot: UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- AB Sciex, LLC
- Distributor
- US Nationwide.
- Lot
- UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABREXAFEMME (Ibrexafungerp) tablets 150 mg, 4 count cartons, Rx Only, Manufactured for and distributed byOctober 18, 2023
- FDAGuider Softip Guide Catheter 5F PRE-SHAPED 40 90CM, Catalog number M003100620, cardiac guide catheterOctober 11, 2023
- FDAGS0008 2Y - Burn Pack - STL, Item Number 880082025October 11, 2023
- FDAArrow ErgoPack Pressure Injectable Arrowg+ard Blue Advance Three-Lumen PICC pre-loaded with Arrow VPSOctober 11, 2023
- FDAIncisive CT, software version 5.0October 11, 2023
- FDASweet & Spicy BBQ sauce, Fresh Chile BBQ sweet & Spicy Hot. 16 fluid ounce glass jar 19 fl oz. (510g) , with aOctober 11, 2023
Questions about this recall
Was Citrine Triple Quad Mass Spectrometer, Part Number 5063684 recalled?
Yes. FDA published recall Z-0040-2024 on October 11, 2023. Reported hazard: In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
What should I do if I have this product?
Voluntary: Firm initiated · Letter