Citrine Triple Quad Mass Spectrometer, Part Number 5063684
Published October 11, 2023 · Recall initiated: August 30, 2023
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- AB Sciex, LLC
- Distributor
- US Nationwide.
- Lot
- UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States