STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.
Published October 30, 2013 · Recall initiated: September 18, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Advanced Sterilization Products
- Distributor
- Nationwide Distribution
- Lot
- Lot# 13F050, 13F051, 13F049
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States