THERMOCOOL SMARTTOUCH SF Uni-Directional Product Codes D-1347-01-S, D-1347-02-S, D-1347-03-S
Published November 8, 2017 · Recall initiated: September 6, 2017
Summary
FDA records medical device Class II recall Z-0054-2018, published November 8, 2017 and initiated September 6, 2017. Recalling firm: Biosense Webster, Inc.; Status: Terminated; Reported quantity: 357 units; Lot: 17682851L 17682730L 17685880L 17685740L 17685692L 17680131L 1768581 lL 17685897L 17685789L 17676998L 17685889L 17685896L 17685792L 17685741L 17685758L 17685846L 17689428L 17689420L 17685881L 17680081L 17689454L 17682788L 17685853L 17685750L 17692068L 17689430L 17680080L 17682787L 17692069L 17689463L 17689421L 17685691L. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biosense Webster, Inc.
- Distributor
- US, Japan, Canada, Australia, New Zealand, Brazil, Croatia, Korea, Austria, Colombia, Italy, Portugal, Argentina, South Africa, Belgium, France, Germany, Ireland, Poland, Netherlands, Sweden, Denmark, Finland, Switzerland, Great Britain, Hungary, Czech Republic, China
- Lot
- 17682851L 17682730L 17685880L 17685740L 17685692L 17680131L 1768581 lL 17685897L 17685789L 17676998L 17685889L 17685896L 17685792L 17685741L 17685758L 17685846L 17689428L 17689420L 17685881L 17680081L 17689454L 17682788L 17685853L 17685750L 17692068L 17689430L 17680080L 17682787L 17692069L 17689463L 17689421L 17685691L
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACARTO VIZIGO Bi-Directional Guiding Sheath, REF D128502 The CARTO VIZIGOTM Bi-Directional Guiding Sheath isMarch 20, 2024
- FDABiosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (Large) C3, REF: D138503, Rx Only, CEMarch 31, 2021
- FDAWaveCrest Left Atrial Appendage (LAA), Occlusion System (32 mm), Catalog Numbers WCR1523 and WCR2523.September 5, 2018
- FDASMARTABLATE RF System GeneratorDecember 13, 2017
- FDAAlcon Custom PakNovember 8, 2017
- FDATHERMOCOOL SF NAV Catheters Model Numbers - BNI35FJCT, BNI35DFCT, D-131503-S, D-131504-SNovember 8, 2017
Questions about this recall
Was THERMOCOOL SMARTTOUCH SF Uni-Directional Product Codes D-1347-01-S, D-1347-02-S, D-1347-03-S recalled?
Yes. FDA published recall Z-0054-2018 on November 8, 2017. Reported hazard: Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.
What should I do if I have this product?
Voluntary: Firm initiated · Letter