Back to search
FDAFDA · Food and drugs·Recall number Z-0055-2021

Quick Release Gait Belt, 60

Published October 14, 2020 · Recall initiated: September 15, 2020

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The buckle on the Gait Belt may break due to degradation of the material due to exposure to gamma radiation.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Medline Industries Inc
Distributor
Domestic Distribution Only: IL, MI
Lot
Model: MDT821208NATM /503689 Lots: 33019040001, 33019090001 and 33019100001 Used in kits with model number: DYKF1045 Lots: 20EBB832 and 20EBM643.
Risk classification
Class II
Area (FDA)
device
Country
United States