FDAFDA · Food and drugs·Recall number Z-0055-2021
Quick Release Gait Belt, 60
Published October 14, 2020 · Recall initiated: September 15, 2020
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The buckle on the Gait Belt may break due to degradation of the material due to exposure to gamma radiation.
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- Medline Industries Inc
- Distributor
- Domestic Distribution Only: IL, MI
- Lot
- Model: MDT821208NATM /503689 Lots: 33019040001, 33019090001 and 33019100001 Used in kits with model number: DYKF1045 Lots: 20EBB832 and 20EBM643.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.