FDAFDA · Food and drugs·Recall number Z-0055-2025

Enterprise Imaging XERO Viewer- software application used for reference and diagnostic viewing of multispecialty medical imaging and non-imaging data . Enterprise Imaging XERO Viewer 8.0.0, 8.0.1, 8.1, 8.1.1, 8.1.2, 8.1.3,8.1.4, 8.2.0.000, 8.2.1.000, 8.2.2.000, 8.3.0.000, 8.3.1.000, 8.3.2.000, 8.4.0.000, 8.4.1.000 (including all SU s and HF s)

Published October 16, 2024 · Recall initiated: September 6, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Software defect, when an image has non-square pixel spacing, the reference/localizer line is incorrectly placed on the scout image in the XERO Viewet, may display the incorrect crosshair/line position on the non-coplanar/scout image with respect to the axial slice location. Inaccurate placement of reference lines or crosshairs have potential risk of misdiagnosis or mistreatment due to unintended shifts in the display of anatomical locations.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Agfa Healthcare NV
Distributor
US Nationwide distribution. Foreign: Argentina, Australia, Belgium, Chile, Canada, Colombia, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Georgia, Greece, Hong Kong, Iceland, Ireland, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Luxembourg, Mexico, Netherlands, Nicaragua, Norway, Oman, Pakistan, Peru, Poland, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Tanzania, Tunisia, United Arab Emirates, United Kingdom
Lot
UDI-DI: 05400874000710 Software Versions affected: All versions of Enterprise Imaging XERO Viewer are affected 8.1.2 or higher 8.1.4 or higher 8.2.0.100 or higher 8.2.2.000 or higher 8.3.2.000 or higher 8.4.0.000 or higher Enterprise Imaging XERO Viewer main version streams 8.4.0, 8.4.1, 8.3.2, 8.3.1, 8.3.0, 8.2.2, 8.2.1, 8.2.0, 8.1.5, 8.1.4, 8.1.3, 8.1.2, 8.1.1, 8.1, 8.0.1, 8.0.0, ICIS View 2014.1, ICIS View 2014.2
Risk classification
Class II
Area (FDA)
device
Country
Belgium