Back to search
FDAFDA · Food and drugs·Recall number Z-0056-2015

Black Mamba Suture Passer. The Biomet Sports Medicine Mamba instruments are utilized to aid in passing suture through soft tissue. The Mamba Needle is designed for use with the Mamba Suture Passer. The needles are sterile, disposable, and for single-patient use only.

Published October 15, 2014 · Recall initiated: September 26, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

An investigation identified that high level friction may exist between the needle and Mamba suture passer instrument, causing the needle to break during use.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Biomet, Inc.
Distributor
Nationwide Distribution-including the states of FL, NY, TX, VA, NV, MI, CA, IN, SD, GA, KY, NC, and AR.
Lot
Catalog Number: 110010849 Lot Number: 169620, 253190
Risk classification
Class II
Area (FDA)
device
Country
United States