Back to search
FDAFDA · Food and drugs·Recall number Z-0057-2013

Hemorrhoidal Circular Stapler and Accessories 33mm Product Usage: The PROXIMATE HCS Hemorrhoidal Circular Staplers, (PPH01 and PPH03), and accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculo mucosal tissue. The TRANSTARTM CIRCULAR STAPLER, (STR10), and accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculomucosal tissue.

Published October 24, 2012 · Recall initiated: August 3, 2012

Summary

FDA records medical device Class I recall Z-0057-2013, published October 24, 2012 and initiated August 3, 2012. Recalling firm: Ethicon Endo-Surgery Inc; Status: Terminated; Reported quantity: 157,881 devices; Lot: PPH01: 68 lots; 19,030 pieces; Expiration date April 2016 -June 2017; PPH03: 494 lots; 144,963 pieces; Expiration date March 2016 - June 2017; STR10: 3 lots; 432 pieces; Expiration date Sept 2016 - Oct 2016. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Ethicon Endo-Surgery initiated this voluntary global recall for specific production lots of PROXIMATE® PPH Hemorrhoidal Circular Stapler and Accessories 33mm (PPH01 and PPH03) and Transtar® Circular Stapler Procedure Set (STR10) due to difficulty firing the device which may result in incomplete firing stroke that may result in an incomplete staple formation.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ethicon Endo-Surgery Inc
Distributor
Worldwide Distribution - US (nationwide) in the states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of: ARGENTINA, AUSTRALIA, BAHRAIN, BELGIUM, BRAZIL, CHILE, CHINA, COLOMBIA, COSTA RICA, CYPRUS, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, GREECE, GUATEMALA, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, REPUBLIC OF KOREA, KUWAIT, LEBANON, MALAYSIA, MEXICO, NEPAL, NICARAGUA, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, PUERTO RICO, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA, VIETNAM, and YEMEN. There was one Canadian consignee and 43 U.S. Government consignees.
Lot
PPH01: 68 lots; 19,030 pieces; Expiration date April 2016 -June 2017; PPH03: 494 lots; 144,963 pieces; Expiration date March 2016 - June 2017; STR10: 3 lots; 432 pieces; Expiration date Sept 2016 - Oct 2016
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Hemorrhoidal Circular Stapler and Accessories 33mm Product Usage: The PROXIMATE HCS Hemorrhoidal Circular Staplers, (PPH01 and PPH03), and accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculo mucosal tissue. The TRANSTARTM CIRCULAR STAPLER, (STR10), and accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculomucosal tissue. recalled?

Yes. FDA published recall Z-0057-2013 on October 24, 2012. Reported hazard: Ethicon Endo-Surgery initiated this voluntary global recall for specific production lots of PROXIMATE® PPH Hemorrhoidal Circular Stapler and Accessories 33mm (PPH01 and PPH03) and Transtar® Circular Stapler Procedure Set (STR10) due to difficulty firing the device which may result in incomplete firing stroke that may result in an incomplete staple formation.

What should I do if I have this product?

Voluntary: Firm initiated · Letter