Mamba Disposable Nitinol Needle. The Biomet Sports Medicine Mamba instruments are utilized to aid in passing suture through soft tissue. The Mamba Needle is designed for use with the Mamba Suture Passer. The needles are sterile, disposable, and for single-patient use only.

Published October 15, 2014 · Recall initiated: September 26, 2014

Summary

FDA records medical device Class II recall Z-0058-2015, published October 15, 2014 and initiated September 26, 2014. Recalling firm: Biomet, Inc.; Status: Terminated; Reported quantity: 164 units; Lot: Catalog Number: 110010851 Lot Number: 139800, 139810, 139820. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

An investigation identified that high level friction may exist between the needle and Mamba suture passer instrument, causing the needle to break during use.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Biomet, Inc.
Distributor
Nationwide Distribution-including the states of FL, NY, TX, VA, NV, MI, CA, IN, SD, GA, KY, NC, and AR.
Lot
Catalog Number: 110010851 Lot Number: 139800, 139810, 139820
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Mamba Disposable Nitinol Needle. The Biomet Sports Medicine Mamba instruments are utilized to aid in passing suture through soft tissue. The Mamba Needle is designed for use with the Mamba Suture Passer. The needles are sterile, disposable, and for single-patient use only. recalled?

Yes. FDA published recall Z-0058-2015 on October 15, 2014. Reported hazard: An investigation identified that high level friction may exist between the needle and Mamba suture passer instrument, causing the needle to break during use.

What should I do if I have this product?

Voluntary: Firm initiated · Letter