Back to search
FDAFDA · Food and drugs·Recall number Z-0064-2015

Omni Micro-electrode/reference electrode for cobas b221 analyzer. The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.

Published October 22, 2014 · Recall initiated: September 12, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The default references for normal values are inconsistent between cobas b 221 and other blood gas instruments. In addition, the reference ranges are different when comparing with corresponding ranges listed in the Instructions for Use for the analyzer and the reference literature source.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Roche Diagnostics Operations, Inc.
Distributor
Nationwide Distribution including Puerto Rico and Guam, and the states of AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, ME, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA and WI.
Lot
Model/Catalog/Part Number: 1. 03337103001 2. 03337111001 3. 03337154001 4. 03337138001 5. 03337120001 6. 03337146001 7. 03337111692 8. 03337138692 9. 03337154692
Risk classification
Class III
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card