FDAFDA · Food and drugs·Recall number Z-0072-2017

ADVIA Centaur XPT Immunoassay System

Published October 19, 2016 · Recall initiated: April 28, 2016

Summary

FDA records medical device Class II recall Z-0072-2017, published October 19, 2016 and initiated April 28, 2016. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: 292 units Total (3 domestically & 289 internationally); Lot: None. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The ADVIA Centaur® XPT default setting for the Daily Maintenance Task (Daily Cleaning Procedure) frequency may have the Daily Maintenance Task frequency set to, As needed instead of Daily. Not performing the daily maintenance may impact any assay and can be detected through monitoring of quality control and calibration results. Systems running in a language other than English are not impacted.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
Worldwide Distribution - US including MD, MN, & MS; **Internationally** AE, AT, AU, BE, BR, CA, CN, DE, DK, EG, ES, FI, FR, GB, GP, HU, IN, IT, JO, JP, KR, MY, NO, NZ, PH, PT, QA, SA, SE, SG, TR & VN
Lot
None
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ADVIA Centaur XPT Immunoassay System recalled?

Yes. FDA published recall Z-0072-2017 on October 19, 2016. Reported hazard: The ADVIA Centaur® XPT default setting for the Daily Maintenance Task (Daily Cleaning Procedure) frequency may have the Daily Maintenance Task frequency set to, As needed instead of Daily. Not performing the daily maintenance may impact any assay and can be detected through monitoring of quality control and calibration results. Systems running in a language other than English are not impacted.

What should I do if I have this product?

Voluntary: Firm initiated · Letter