Sunquest Laboratory : intended for use by professionals working in a clinical laboratory.

Published October 24, 2012 · Recall initiated: November 15, 2011

Summary

FDA records medical device Class III recall Z-0073-2013, published October 24, 2012 and initiated November 15, 2011. Recalling firm: Sunquest Information Systems, Inc.; Status: Terminated; Reported quantity: 127 sites; Lot: Versions 6.4.0 and later. Class III means the product is not likely to cause adverse health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Online calculations may not be performed as expected or may yield results that are different than expected when the calculation module is called from the Online Entry (OEM) application for a subtest whose subcolumn number on the instrument online string is divisible by 10.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Sunquest Information Systems, Inc.
Distributor
Nationwide distribution
Lot
Versions 6.4.0 and later
Risk classification
Class III
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Sunquest Laboratory : intended for use by professionals working in a clinical laboratory. recalled?

Yes. FDA published recall Z-0073-2013 on October 24, 2012. Reported hazard: Online calculations may not be performed as expected or may yield results that are different than expected when the calculation module is called from the Online Entry (OEM) application for a subtest whose subcolumn number on the instrument online string is divisible by 10.

What should I do if I have this product?

Voluntary: Firm initiated · Letter