Sunquest Laboratory : intended for use by professionals working in a clinical laboratory.
Published October 24, 2012 · Recall initiated: November 15, 2011
Summary
FDA records medical device Class III recall Z-0073-2013, published October 24, 2012 and initiated November 15, 2011. Recalling firm: Sunquest Information Systems, Inc.; Status: Terminated; Reported quantity: 127 sites; Lot: Versions 6.4.0 and later. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Online calculations may not be performed as expected or may yield results that are different than expected when the calculation module is called from the Online Entry (OEM) application for a subtest whose subcolumn number on the instrument online string is divisible by 10.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Sunquest Information Systems, Inc.
- Distributor
- Nationwide distribution
- Lot
- Versions 6.4.0 and later
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASunquest Laboratory, version 7.0 not availableNovember 14, 2012
- FDASunquest Application Interfacing Outbound Phlebotomy Point of Care Interface, v3.0.0_07222004 and later.November 14, 2012
- FDA***This recall is being conducted due to an incomplete recall for RES 57829*** Curlin Infusion AdministrationOctober 24, 2012
- FDA"***Electric Systems Foot Control****Rx Only***" Product Usage: Cutting and shaping bone including bones ofOctober 24, 2012
- FDASunquest Laboratory Sunquest Laboratory is intended for use by professionals working in a clinical laboratory.September 19, 2012
- FDASunquest Laboratory LabAccess Results Workstation (LARS)September 5, 2012
Questions about this recall
Was Sunquest Laboratory : intended for use by professionals working in a clinical laboratory. recalled?
Yes. FDA published recall Z-0073-2013 on October 24, 2012. Reported hazard: Online calculations may not be performed as expected or may yield results that are different than expected when the calculation module is called from the Online Entry (OEM) application for a subtest whose subcolumn number on the instrument online string is divisible by 10.
What should I do if I have this product?
Voluntary: Firm initiated · Letter