Back to search
FDAFDA · Food and drugs·Recall number Z-0073-2013

Sunquest Laboratory : intended for use by professionals working in a clinical laboratory.

Published October 24, 2012 · Recall initiated: November 15, 2011

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Online calculations may not be performed as expected or may yield results that are different than expected when the calculation module is called from the Online Entry (OEM) application for a subtest whose subcolumn number on the instrument online string is divisible by 10.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Sunquest Information Systems, Inc.
Distributor
Nationwide distribution
Lot
Versions 6.4.0 and later
Risk classification
Class III
Area (FDA)
device
Country
United States