FDAFDA · Food and drugs·Recall number Z-0073-2014
NexGen¿ Femoral Component (CR-FLEX GSF PRECOAT SZ C-L; CR-FLEX GSF PRECOAT SZ C-R ; CR-FLEX GSF PCT SZ C-L MINUS; CR-FLEX GSF PCT SZ C-R MINUS ;CR-FLEX GSF PRECOAT SZ D-L; CR-FLEX GSF PRECOAT SZ D-R ;CR-FLEX GSF PCT SZ D-L MINUS ; CR-FLEX GSF PCT SZ D-R MINUS ;CR-FLEX GSF PRECOAT SZ E-L CR-FLEX GSF PRECOAT SZ E-R ;CR-FLEX GSF PCT SZ E-L MINUS ;CR-FLEX GSF PCT SZ E-R MINUS ;CR-FLEX GSF PRECOAT SZ F-L ;CR-FLEX GSF PRECOAT SZ F-R ;CR-FLEX GSF PCT SZ F-L MINUS ;CR-FLEX GSF PCT SZ F-R MINUS ;CR-FLEX GSF PRECOAT SZ G-L ;CR-FLEX GSK PRECOAT SZ G-R ;CR-FLEX GSF PCT SZ G-L MINUS ;CR-FLEX GSF PCT SZ G-R MINUS ; LPS-FLEX GSF OPT SZ C-L ; LPS-FLEX GSF OPT SZ C-R ;LPS-FLEX GSF OPT SZ D-L ;LPS-FLEX GSF OPT SZ D-R ;LPS-FLEX GSF OPT SZ E-L ;LPS-FLEX GSF OPT SZ E-R ;LPS-FLEX GSK OPT SZ F-L ;LPS-FLEX GSF OPT SZ F-R ;LPS-FLEX GSF OPT SZ G-L ; LPS-FLEX GSF OPT SZ G-R ;CR-FLEX PCT FEM B-L; CR-FLEX PCT FEM B-R; CR-FLEX PCT FEM C-L; CR-FLEX PCT FEM C-R; CR-FLEX PCT FEM C-L MINUS; CR-FLEX PCT FEM C-R MINUS; CR-FLEX PCT FEM D-L; CR-FLEX PCT FEM D-R; CR-FLEX PCT FEM D-L MINUS; CR-FLEX PCT FEM D-R MINUS; CR-FLEX PCT FEM E-L; CR-FLEX PCT FEM E-R; CR-FLEX PCT FEM E-L MINUS; CR-FLEX PCT FEM E-R MINUS; CR-FLEX PCT FEM F-L; CR-FLEX PCT FEM F-R; CR-FLEX PCT FEM F-L MINUS; CR-FLEX PCT FEM F-R MINUS; CR-FLEX PCT FEM G-L; CR-FLEX PCT FEM G-R; CR-FLEX PCT FEM G-L MINUS; CR-FLEX PCT FEM G-R MINUS; CR-FLEX POR FEM B-L; CR-FLEX POR FEM B-R; CR-FLEX POR FEM C-L; CR-FLEX POR FEM C-R; CR-FLEX POR FEM C-L MINUS; CR-FLEX POR FEM C-R MINUS; CR-FLEX POR FEM D-L; CR-FLEX POR FEM D-R; CR-FLEX POR FEM D-L MINUS; CR-FLEX POR FEM D-R MINUS; CR-FLEX POR FEM E-L; CR-FLEX POR FEM E-R; CR-FLEX POR FEM E-L MINUS; CR-FLEX POR FEM E-R; MINUS; CR-FLEX POR FEM F-L; CR-FLEX POR FEM F-R ;CR-FLEX POR FEM F-L MINUS; CR-FLEX POR FEM F-R MINUS; CR-FLEX POR FEM G-L; CR-FLEX POR FEM G-R; CR-FLEX POR FEM G-L MINUS CR-FLEX POR FEM G-R MINUS; CR-FLEX OPT FEM B-L; CR-FLEX OPT FEM B-R; CR-FLEX OPT FEM C-L; CR-FLEX OPT FEM C-R; CR-FLEX OPT FEM C-L MINUS; CR-FLEX OPT FEM C-R MINUS; CR-FLEX OPT FEM D-L; CR-FLEX OPT FEM D-R; CR-FLEX OPT FEM D-L MINUS; CR-FLEX OPT FEM D-R MINUS; CR-FLEX OPT FEM E-L; CR-FLEX OPT FEM E-R; CR-FLEX OPT FEM E-L MINUS; CR-FLEX OPT FEM E-R MINUS; CR-FLEX OPT FEM F-L; CR-FLEX OPT FEM F-R; CR-FLEX OPT FEM F-L MINUS; CR-FLEX OPT FEM F-R MINUS; CR-FLEX OPT FEM G-L; CR-FLEX OPT FEM G-R; CR-FLEX OPT FEM G-L MINUS; CR-FLEX OPT FEM G-R MINUS; LPS-FLEX CO-NID FEM A-L ; LPS-FLEX CO-NID FEM A-R ; LPS-FLEX RECOAT FEMORAL A-L; LPS-FLEX PRECOAT FEMORAL A-R; LPS-FLEX CO-NID FEM B-L ; LPS-FLEX CO-NID FEM B-R ; LPS-FLEX PRECOAT FEMORAL B-L; LPS-FLEX PRECOAT FEMORL B-R; LPS-FLEX CO-NID FEM C-L ; LPS-FLEX CO-NID FEM C-R ; LPS-FLEX PRECOAT FEMORAL C-L; LPS-FLEX PRECOAT FEMORAL C-R; LPS-FLEX CO-NID FEM D-L ; LPS-FLEX CO-NID FEM D-R ; LPS-FLEX PRECOAT FEMORAL D-L; LPS-FLEX PRECOAT FEMORAL D-R; LPS-FLEX CO-NID FEM E-L ; LPS-FLEX CO-NID FEM E-R ; LPS-FLEX PRECOAT FEMORAL E-L; LPS-FLEX PRECOAT FEMORAL E-R; LPS-FLEX CO-NID FEM F-L ; LPS-FLEX CO-NID FEM F-R ; LPS-FLEX PRECOAT FEMORAL F-L; LPS-FLEX PRECOAT FEMORAL F-R; LPS-FLEX CO-NID FEM G-L ; LPS-FLEX CO-NID FEM G-R ; LPS-FLEX PRECOAT FEMORAL G-L; LPS-FLEX PRECOAT FEMORAL G-R; LPS-FLEX PRECOAT FEMORAL H-L; LPS-FLEX PRECOAT FEMORAL H-R ; LPS-FLEX POROUS FEMORAL B-L ; LPS-FLEX POROUS FEMORAL B-R ; LPS-FLEX POROUS FEMORAL C-L ; LPS-FLEX POROUS FEMORAL C-R LPS-FLEX POROUS FEMORAL D-L ; LPS-FLEX POROUS FEMORAL D-R ; LPS-FLEX POROUS FEMORAL E-L ; LPS-FLEX POROUS FEMORAL E-R ; LPS-FLEX POROUS FEMORAL F-L ; LPS-FLEX POROUS FEMORAL F-R ;LPS-FLEX POROUS FEMORAL G-L ; LPS-FLEX POROUS FEMORAL G-R ; LPS-FLEX POROUS FEMORAL H-L ; LPS-FLEX POROUS FEMORAL H-R ; LPS-FLEX OPTION FEMORAL A-L; LPS-FLEX OPTION FEMORAL A-R; LPS-FLEX OPTION FEMORAL B-L; LPS-FLEX OPTION FEMORAL B-R; LPS-FLEX OPTION FEMORAL B-L ; LPS-FLEX OPTION FEMORAL B-R ; LPS-FLEX OPTION FEMORAL C-L; LPS-FLEX OPTION FEMORAL C-R; LPS-FLEX OPTION FEMORAL C-L; LPS-FLEX OPTION FEMORAL C-R; LPS-FLEX OPTI
Published November 6, 2013 · Recall initiated: August 30, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Honduras, Hong Kong, India, Israel, Jamaica, Japan, Korea, Malaysia, Mexico, New Zealand, Nicaragua, Panama, Peru, Singapore, Taiwan, Thailand, and Venezula.
- Lot
- Part Number Range: 005750, 5764, 5950, 5952, 5956, 5960, 5962, 5964, 5966, 5970, 5972, 5990, 5992, 5994, and 5996-011/018-01/92.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States