Back to search
FDAFDA · Food and drugs·Recall number Z-0082-2014

MAQUET Getinge Group CARDIOHELP Base Unit REF 70104.7999 REF 70104.8012 Manufactured for: MAQUET Cardiopulmonary AG Made in Germany. Cardiopulmonary support system.

Published November 6, 2013 · Recall initiated: September 12, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

It has come to the attention of MAQUET that in certain rare instances, the CARDIOHELP-i device may display an error message with audible alarm, indicating "battery defective." The alarm will occur-upon start up or when the main power supply is unplugged from the main power source. The human machine interface touch screen of the CARDIOHELP-i device may gradually become black due to the human machine interface (HMI) back light shutting down.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Maquet Medical Systems USA
Distributor
Nationwide Distribution including AZ, FL, CO, TX, OH, VA, MN, WI, SC, IL, ND, UT, CA, OR, NY, IA, CT, SD, MI, HI, AL, MD, KS, MO
Lot
Part No. 70104.8012 Serial Nos. 90410440; 90410452; 90410455; 90410456; 90410457; 90410459; 90410469; 90410474; 90410480; 904104781; 90410482; 90410484; 90410498; 90410499; 90410500; 90410501; 90410502; 90410503; 90410521; 90410522; 90410524; 90410526; 90410527; 90410531; 90410533; 90410534; 90410535; 90410550; 90410551; 90410552; 90410568; 90410571; 90410572; 90410573; 90410575; 90410576; 90410577; 90410580; 90410581; 90410582; 90410584; 90410585; 90410592; 90410596; 90410599; 90410600; 90410605; 90410606; 90410607; 90410608; 90410609; 90410610; 90410611; 90410612; 90410626; 90410638; 90410640; 90410642; 90410644; 90410646; 90410647; 90410650; 90410651; 90410652; 90410653; 90410656; 90410659; 90410660; 90410661; 90410662; 90410663; 90410664; 90410665; 90410666; 90410667; 90410668; 90410670; 90410700; 90410708; 90410846; 90410750; 90410786
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card