Back to search
FDAFDA · Food and drugs·Recall number Z-0089-2023

Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A

Published October 26, 2022 · Recall initiated: August 24, 2022

Summary

FDA records medical device Class II recall Z-0089-2023, published October 26, 2022 and initiated August 24, 2022. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Ongoing; Reported quantity: 1592 units; Lot: UDI-DI : 0084276802043821250BA2022090710464340840 Lot Number: 21250BA Exp. Date: 2022-09-07. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Dimension Vista® High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result ranging from -90% to 133% may be observed.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
Nationwide Foreign: Canada
Lot
UDI-DI : 0084276802043821250BA2022090710464340840 Lot Number: 21250BA Exp. Date: 2022-09-07
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A recalled?

Yes. FDA published recall Z-0089-2023 on October 26, 2022. Reported hazard: Dimension Vista® High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result ranging from -90% to 133% may be observed.

What should I do if I have this product?

Voluntary: Firm initiated · Letter