FDAFDA · Food and drugs·Recall number Z-0092-2014
REF 5100-088, TPS Micro Driver, For Use with Stryker Total Performance System Caution Federal Law (USA) restricts the device to use by or on the order of a physician Driver, Wire and Bone Drill; The Stryker Total Performance System (TPS) is intended for use in the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues for a variety of applications such as ENT, dental, orthopedic, maxillofacial, spinal, and plastic surgery. The instruments are also used in the placement of screws, wires, pins, and other fixation devices.
Published November 6, 2013 · Recall initiated: October 2, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was discovered that the silicone potting failed to properly cure after undergoing the prescribed curing possess. The potting material is used to insulate, stabilize and provide a moisture barrier to the TPS MicroDriver main Printed Circuit Board Assembly (PCBA). Inadequate cure of the silicone potting could potentially allow it to leak to the outside of the handpiece or cause an electrical ma
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Stryker Instruments Div. of Stryker Corporation
- Distributor
- Worldwide distribution: US (nationwide) and countries of: Australia, Canada, France, Germany, Italy, Japan, Spain, Switzerland, and United Kingdom.
- Lot
- 10933, 11433, 11483, 20553, 30163, 31553, 40713, 40833, 41113, 50463, 60133, 60423, 90383, 110093, 110193, 1030133, 1030403, 1040413, 2010013, 2010123, 96090403, 96091343, 97020193, 97030013, 97030543, 97050243, 97060163, 97080033, 98010083, 98020703, 98030103, 98040213, 98060023, 98061033, 98070603, 98090023, 98090473, 98090503, 98101143, 98110323, 99010433, 99030253, 99050033, 99050383, 99060163, 99060423, 99061063, 99061283, 99070453, 99071293, 99080243, 99080483, 99090913, 99101273, 221303733, 326102033, 434909453, 500800103, 504900233, 631112793, 631414853, 700314923, 723606673, 725315933, 735502183, 813304543, and 921700523 ,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States