FDAFDA · Food and drugs·Recall number Z-0099-2017

URISYS 2400 Urine Analyzer Automated Urinalysis System Product Usage: The RSD 800 is a computer-controlled stand-alone sorting system for sorting barcoded samples. The system has an online connection to the host.

Published November 2, 2016 · Recall initiated: July 15, 2016

Summary

FDA records medical device Class II recall Z-0099-2017, published November 2, 2016 and initiated July 15, 2016. Recalling firm: Roche Diagnostics Operations, Inc.; Status: Terminated; Reported quantity: 9014 in total; Lot: Not applicable. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Roche Diagnostics Operations, Inc.
Distributor
US Nationwide Distribution
Lot
Not applicable
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was URISYS 2400 Urine Analyzer Automated Urinalysis System Product Usage: The RSD 800 is a computer-controlled stand-alone sorting system for sorting barcoded samples. The system has an online connection to the host. recalled?

Yes. FDA published recall Z-0099-2017 on November 2, 2016. Reported hazard: Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

What should I do if I have this product?

Voluntary: Firm initiated · Letter