Back to search
FDAFDA · Food and drugs·Recall number Z-0107-2027

SEDIPLAST; Model/Catalog Numbers: S-100 and S-1000;

Published October 14, 2026 · Recall initiated: August 25, 2026

Summary

FDA records medical device Class II recall Z-0107-2027, published October 14, 2026 and initiated August 25, 2026. Recalling firm: POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC; Status: Ongoing; Reported quantity: 448400; Lot: Model/Catalog No. S-100 and S-1000; UDI-DI: N/A; Lot Number E1150A; Expiration Date 06/30/27. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the tube cap to detach or be forcibly ejected, potentially resulting in the release of blood and exposure of the operator or surrounding work area to a biological hazard (blood).

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC
Distributor
Worldwide - US Nationwide distribution in the states and territory of AK, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV and the country of United Kingdom.
Lot
Model/Catalog No. S-100 and S-1000; UDI-DI: N/A; Lot Number E1150A; Expiration Date 06/30/27
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SEDIPLAST; Model/Catalog Numbers: S-100 and S-1000; recalled?

Yes. FDA published recall Z-0107-2027 on October 14, 2026. Reported hazard: During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the tube cap to detach or be forcibly ejected, potentially resulting in the release of blood and exposure of the operator or surrounding work area to a biological hazard (blood).

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail