St Jude Medical, 10F Fast-Cath Trio" Hemostasis Introducer, REF # 406308, Sterile EO. The Fast-Cath Trio" Hemostasis Introducer consists of the following components: Port hemostasis adapter (3), Sheath (10F Cath-Lock"), dilator (10F) and 0.038 guidewire. See product labeling for diagram of contents.
Published October 29, 2014 · Recall initiated: October 14, 2014
Summary
FDA records medical device Class II recall Z-0108-2015, published October 29, 2014 and initiated October 14, 2014. Recalling firm: St Jude Medical; Status: Terminated; Reported quantity: 220; Lot: Batch 4641580. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
St. Jude Medical is conducting a voluntary recall of SMJ Batch # 4641580 of the Fast-Cath Trio" Hemostasis Introducer, Catalog number 406308. The inner diameter of the 10F dilator is too small, such that the guidewire cannot advance through the device which may result in a procedural delay while another dilator of the desired size is obtained.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- St Jude Medical
- Distributor
- Nationwide Distribution including FL, IN, ME, MA, NY, NY, OH, RI,TX, and WV.
- Lot
- Batch 4641580
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASt. Jude Medical, Fixed Curve Response 6F Electrophysiology Catheter, CSL, Model 401353November 12, 2014
- FDAMindray V21, Size: 515 mm X 335 mm X 685 mm, N.W.: 8 kg, G.W.: 11 kg, Qty:1 The V-Series Monitor is intendedOctober 29, 2014
- FDAMicroseal Augmented (20 degrees) Liner, acetabular liner component made from UHMW polyethylene (ASTM F648).October 29, 2014
- FDABox Label: 3 mm Cueva Electrode Kit, Sterile EO, Rx Only, REF. CNE1-3KDINX . Pouch labels: Each kit contains 3October 29, 2014
- FDAMortara Surveyor S12 and Surveyor S19 Patient Monitors. Product Usage: They are indicated for use in adultOctober 29, 2014
- FDAPhilips Avalon Monitors with software revision J.30.58: Model Product FM20 M2702A ; FM30 M2703A ; FM50 M2705AOctober 29, 2014
Questions about this recall
Was St Jude Medical, 10F Fast-Cath Trio" Hemostasis Introducer, REF # 406308, Sterile EO. The Fast-Cath Trio" Hemostasis Introducer consists of the following components: Port hemostasis adapter (3), Sheath (10F Cath-Lock"), dilator (10F) and 0.038 guidewire. See product labeling for diagram of contents. recalled?
Yes. FDA published recall Z-0108-2015 on October 29, 2014. Reported hazard: St. Jude Medical is conducting a voluntary recall of SMJ Batch # 4641580 of the Fast-Cath Trio" Hemostasis Introducer, Catalog number 406308. The inner diameter of the 10F dilator is too small, such that the guidewire cannot advance through the device which may result in a procedural delay while another dilator of the desired size is obtained.
What should I do if I have this product?
Voluntary: Firm initiated · Letter