Back to search
FDAFDA · Food and drugs·Recall number Z-0114-2018

3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000)

Published November 29, 2017 · Recall initiated: October 17, 2017

Summary

FDA records medical device Class II recall Z-0114-2018, published November 29, 2017 and initiated October 17, 2017. Recalling firm: 3M Company - Health Care Business; Status: Terminated; Reported quantity: 153 cases (3825 sensors); Lot: Lot Numbers: 17230, Exp. 2020 Aug 18; 17233, Exp. 2020 Aug 21; 17234, Exp. 2020 Aug 22; 17241, Exp. 2020 Aug 29. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower than the patients actual temperature.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
3M Company - Health Care Business
Distributor
CA, MI, IL, MO, MN, GA, and Canada, Switzerland
Lot
Lot Numbers: 17230, Exp. 2020 Aug 18; 17233, Exp. 2020 Aug 21; 17234, Exp. 2020 Aug 22; 17241, Exp. 2020 Aug 29
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was 3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000) recalled?

Yes. FDA published recall Z-0114-2018 on November 29, 2017. Reported hazard: During a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower than the patients actual temperature.

What should I do if I have this product?

Voluntary: Firm initiated · Letter