Halyard Closed Suction System for Adults, 14 F, Elbow, Directional Tip, Wet Pak. 12 vials per kit, 20 kits per case.
Published November 29, 2017 · Recall initiated: October 6, 2017
Summary
FDA records medical device Class II recall Z-0120-2018, published November 29, 2017 and initiated October 6, 2017. Recalling firm: Halyard Health, Inc; Status: Terminated; Reported quantity: 52,660 kits (2,633 cases) with 631,920 saline vials; Lot: Product Code 22169, Lots AB6123U08, AB6123U11. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Received recall notification from Smiths Medical that lots of 15ml Saline Vials may be susceptible to leaking and could potentially result in the exposure of the saline contents to infectious agents.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Halyard Health, Inc
- Distributor
- Distributed domestically to AL, AZ, CA, CO, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OK, PA, PR, RI, SC, SD, TN, TX, VA, WA, WV
- Lot
- Product Code 22169, Lots AB6123U08, AB6123U11
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHalyard Closed Suction System for Adults with BALLARD Technology WET PAK. 12 saline vials per kit, 20 kits perNovember 29, 2017
- FDAHalyard Closed Suction System for Adults, 14 F, DSE, Wet Pak. 12 or 6 vials per kit, 20 kits per case.November 29, 2017
- FDAHalyard Closed Suction System for Adults, 14 F, T-Piece, Wet Pak. 12 or 6 vials per kit, 20 kits per case.November 29, 2017
- FDAHalyard Turbo-Cleaning Closed Suction System for Adults, 14 F, T-Piece, Wet Pak. 12 vials per kit, 20 kits perNovember 29, 2017
- FDAROSA BRAIN 3.0.0November 29, 2017
- FDANOVAPLUS* Endotracheal WET PAK* Closed Suction System for Adults with BALLARD* Technology, Double SwivelNovember 29, 2017
Questions about this recall
Was Halyard Closed Suction System for Adults, 14 F, Elbow, Directional Tip, Wet Pak. 12 vials per kit, 20 kits per case. recalled?
Yes. FDA published recall Z-0120-2018 on November 29, 2017. Reported hazard: Received recall notification from Smiths Medical that lots of 15ml Saline Vials may be susceptible to leaking and could potentially result in the exposure of the saline contents to infectious agents.
What should I do if I have this product?
Voluntary: Firm initiated · Letter