Yuno Mobile Operating Table - Product Usage: Intended for support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment. Part 143301F0 (US) Part 143301B0 (OUS)
Published October 28, 2020 · Recall initiated: September 24, 2020
Summary
FDA records medical device Class II recall Z-0132-2021, published October 28, 2020 and initiated September 24, 2020. Recalling firm: Maquet Cardiovascular Us Sales, Llc; Status: Terminated; Reported quantity: 62 units US; Lot: Part 143301F0: SN 1 to SN 90 Part 143301B0: SN 1 to SN 267 (OUS). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Plastic socket of the IEC 60320-1 socket pulled out of the metal cage and electronic parts exposed; where the socket was pulled out and main cable is still connected to the main socket (power source), there is a risk that a user or third person may suffer an electric shock when coming in contact with the exposed electronic components
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Maquet Cardiovascular Us Sales, Llc
- Distributor
- Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, BAHRAIN, BOLIVIA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, FINLAND, FRANCE GERMANY, HONG KONG, HUNGARY, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, JAPAN, KENYA, KOREA, KUWAIT, LATVIA, LEBANON, MACAU, MACEDONIA, MALAYSIA, MEXICO, NEW ZEALAND, OMAN, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, RWANDA, SAUDI ARABIA, SERBIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TUNESIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM.
- Lot
- Part 143301F0: SN 1 to SN 90 Part 143301B0: SN 1 to SN 267 (OUS)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHLS Set Advanced, Model: BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Product Code/Part Number: 70105.2794 - ProductDecember 23, 2020
- FDAHLS Set Advanced Model Name: BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code/Part Number: 70105.2797 -December 23, 2020
- FDAMR Coils The MR Coil is intended to be used in conjunction with a Magnetic Resonance Scanner to produceOctober 28, 2020
- FDATWINFIX 5.0MM AB SUTURE ANCHOR WITH TWO Absorbable Suture anchors Product Number: 7210309October 28, 2020
- FDAMaquet/Getinge Rotaflow System Console US (115 V / 60 Hz) - Product Usage: Instructions for Use - CentrifugalSeptember 23, 2020
- FDAAvalon Elite Vascular Access Kit, Model No. 70106.3540, UDI 04037691831039April 8, 2020
Questions about this recall
Was Yuno Mobile Operating Table - Product Usage: Intended for support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment. Part 143301F0 (US) Part 143301B0 (OUS) recalled?
Yes. FDA published recall Z-0132-2021 on October 28, 2020. Reported hazard: Plastic socket of the IEC 60320-1 socket pulled out of the metal cage and electronic parts exposed; where the socket was pulled out and main cable is still connected to the main socket (power source), there is a risk that a user or third person may suffer an electric shock when coming in contact with the exposed electronic components
What should I do if I have this product?
Voluntary: Firm initiated · Letter